52 Remote Programming
SCPI Command Completion
SCPI commands sent to the power supply are processed either sequentially or in parallel. Sequential commands finish
execution before a subsequent command begins. A parallel command can begin execution while a preexisting command is
still executing (overlapping commands). Commands that affect trigger actions are among the parallel commands.
The
*WAI, *OPC
,
and
*OPC?
common commands provide different ways of indicating when all transmitted commands,
including any parallel ones, have completed their operations. The syntax and parameters for these commands are described
in Chapter 7 - "Language Dictionary". Some practical considerations for using these commands are as follows:
*WAI
This prevents the power supply from processing subsequent commands until all pending operations
are completed. If something prevents completion of an existing operation,
*WAI
can place the power
supply and the controller in a "hang- up" condition.
*OPC?
This places a 1 in the Output Queue when all pending operations have completed. Because it
requires your program to read the returned value from the queue before executing the next program
statement,
*OPC?
could prevent subsequent commands from being executed.
*OPC
This sets the OPC status bit when all pending operations have completed. Since your program can
read this status bit on an interrupt basis,
*OPC
allows subsequent commands to be executed.
The trigger subsystem must be in the Idle state in order for the status OPC bit to be true. Therefore, as far as triggers are
concerned, OPC is false whenever the trigger subsystem is in the Initiated state. However, OPC is also false if there are any
commands still pending.
Note
For a detailed discussion of
*WAI, *OPC
and
*OPC?,
see "Device/Controller Synchronization Techniques"
in ANSI/IEEE Std 488.2.
Programming Examples
The examples given here are generic, without regard to the programming language or type of GPIB interface. Because
SCPI commands are sent as ASCII output strings within the programming language statements, the SCPI syntax is
independent of both programming language and interface.
Note
The examples are followed by sample program code written for three popular types of BASIC-controlled
GPIB interfaces.
Important
The power supply responds simultaneously to both digital and analog progamming inputs. If it is
receiving an input over the GPIB and a corresponding input from the front panel (and/or from the analog
programming port), the power supply output will be the algebraic sum of the inputs.