Remote Programming 53
Programming Voltage and Current
The following statements program both voltage and current and return the actual output from the sense terminals:
OUTP OFF
Disable the output.
VOLT 45;CURR 25
Program the voltage and current.
VOLT?;CURR?
Read back the programmed levels.
OUTP ON
Enable the output.
MEAS:VOLT?;MEAS:CURR?
Read back the outputs from the sense terminals.
Programming Protection Circuits
This example programs the voltage and current, programs an overvoltage protection value, and turns on the overcurrent
protection. It then reads back all the programmed values.
VOLT:LEV 45;PROT 47.5
Program the voltage and overvoltage protection.
CURR:LEV 25;PROT:STAT ON
Program the current and overcurrent protection.
VOLT:LEV?;PROT?;:CURR:LEV?;PROT:STAT?
Read back the programmed values.
Note the required use of the optional
LEVel
header in the above example (see "The Effect of Optional Headers", given
previously).
Changing Outputs by Trigger
If you do not program pending triggered levels, they default to the programmed (immediate) output levels. The following
statements shows some basic trigger commands.
OUTP OFF
Disable the output.
VOLT:LEV:IMM 22;TRIG 25
Program the (immediate) voltage level to 22V and the pending
triggered level to 25 V.
CURR:LEV:IMM I5;TRIG 25
Program the (immediate) current level to 15 A and the pending
triggered level to 25 A.
VOLT:LEV:IMM?;TRIG?;:CURR:LEV:IMM?;TRIG?
Check all the programmed values.
OUTP ON
Enable the output.
MEAS:VOLT?;CURR?
Read back the immediate levels from the sense terminals.
INIT;TRIG
Arm the trigger circuit and send a single trigger.
INIT;*TRG
Same as above, except using a common command.
MEAS:VOLT?;CURR?
Read back the triggered levels from the sense terminals.
If you need to send two or more triggers, program the trigger circuit for continuous arming.
OUTP OFF
Disable the output.
VOLT:LEV:IMM 5O;TRIG 25
Program the (immediate) voltage level to 50 V and the pending
triggered level to 25 V.
INTIT:CONT ON
Program the trigger circuit for continuous arming.
OUTP ON
Enable the output to 50 V.
TRIG
Trigger the output voltage to 25 V.
VOLT:TRIG 50;:TRIG
Set the pending trigger level to 50 V and trigger the output
voltage back to 50 V.
INTIT:CONT OFF
Remove the continuous triggger arming.
Содержание E4356A
Страница 1: ...Operating and Programming Guide Keysight Model E4356A Telecommunications DC Power Supply ...
Страница 2: ......
Страница 6: ...5 Declaration Page ...
Страница 13: ......
Страница 109: ......
Страница 113: ......
Страница 119: ......