Calibration 103
BASIC Calibration Program
The following program can be run on any controller operating under Keysight BASIC. The assumed power supply address
is 5 and calibration password is 4356. If required, change these parameters in the appropriate statements.
10
! Keysight BASIC Calibration Program
20
!
30
DIM Resp$ [255],Err_msg$[255]
40
!
50 Volt_cal: ! Voltage DAC calibration
60
Err_found=0
70
PRINT TABXY(5,10),"CONNECT INSTRUMENTS AS SHOWN IN FIG. A-1(1). Then Press Continue"
80
PAUSE
90
CLEAR SCREEN
100
!
110
! Assign power supply GPIB address
120
!
130
ASSIGN @Ps TO 705
140
!
150
! Initialize power supply
160
!
170
OUTPUT @Ps;"*RST;OUTPUT ON"
180
!
190
! Password is optional - only required if set to non-zero value
200
! Default password is the four-digit model number
210
!
220
! LINE 240 PASSWORD MUST BE EDITED FOR MODEL OTHER THAN 6680
230
!
240
OUTPUT @Ps;"CAL:STATE ON, 4356"
250
1
260
OUTPUT @Ps;"CAL:VOLTAGE:LEVEL MIN"
270
INPUT "ENTER VOLTAGE MEASUREMENT FROM EXTERNAL VOLTMETER",Volt_read
280
OUTPUT @Ps;"CAL:VOLTAGE ";Volt_read
290
OUTPUT @Ps;"CAL:VOLTAGE:LEVEL MAX"
300
INPUT "ENTER VOLTAGE MEASUREMENT FROM EXTERNAL VOLTMETER",Volt_read
310
OUTPUT @Ps;"CAL:VOLTAGE ";Volt_read
320
!
330
! Calibrate overvoltage protection circuit
340
!
350
OUTPUT @Ps;"CAL:VOLTAGE:PROTECTION"
360
!
370
GOSUB Save_cal
380
IF Err_found THEN
390
INPUT "ERRORS have occurred, REPEAT VOLTAGE CALIBRATION ( Y 0R N )?",Resp$
400
IF TRIM$(UPC$(Resp$[1,1] ) )="Y" THEN GOTO Volt_cal
410
END IF
420
IF Err_found THEN
430
PRINT "VOLTAGE CALIBRATION NOT SAVED"
440
ELSE
450
PRINT "VOLTAGE CALIBRATION COMPLETE"
460
END IF
470
!