78 Language Dictionary
OUTP:REL
This command is valid only if the power supply is configured for the optional relay connector. Programming
ON
closes the
relay contacts; programming
OFF
opens them. The relay is controlled independently of the output state. If the power
supply is supplying power to a load, that power will appear at the relay contacts during switching. If the power supply is not
configured for the relay option, sending either relay command generates an error.
Command Syntax
OUTPut:RELay[:STATe] <bool>
Parameters
0 | 1 | OFF | ON
*RST Value
0
Examples
OUTP:REL 1 OUTP:REL OFF
Query Syntax
OUTPput:RELay?
Returned Parameters
0 |1
Related Commands
OUTP[:STAT] *RCL *SAV
OUTP:REL:POL
This command is valid only if the power supply is configured for the optional relay connector. Programming
NORMal
causes the relay output polarity to be the same as the power supply output. Programming
REV
erse causes the relay output
polarity to be opposite to that of the power supply output. If
OUTP[:STAT] = ON
when either relay command is sent, the
power supply output voltage is set to 0 during the time that the relays are changing polarity. If the power supply is not
configured for the relay option, sending either relay command generates an error.
Command Syntax
OUTPut:RELay:POLarity <CRD>
Parameters
NORMal | REVerse
*RST Value
NORM
Examples
OUTP:REL:POL NORM
Query Syntax
OUTPput:RELay:POLarity?
Returned Parameters
NORM | REV
Related Commands
OUTP[:STAT] *RCL *SAV
Status Subsystem
This subsystem programs the power supply status registers. The power supply has three groups of status registers;
Operation, Questionable,
and
Standard Event.
The Standard Event group is programmed with Common commands as
described in "Chapter 8 - Status Reporting". The Operation and Questionable status groups each consist of the Condition,
Enable, and Event registers and the NTR and PTR filters.
STAT:PRES
This command sets all defined bits in the Status Subsystem PTR registers and clears all bits in the subsystem NTR and
Enable registers. STAT:OPER:PTR is set to 1313 and STAT:QUES:PTR is set to 1555.
Command Syntax
STATus:PRESet
Parameters
(None)
Examples
STAT:PRES STATUS:PRESET
Query Syntax
(None)
Related Commands
(None)