56
4000 X-Series Oscilloscopes Advanced Training Guide
3
Advanced Triggering, Search & Navigation, and Segmented Acquisition Labs
Your scope’s display should now look similar to
. Using the scope’s
default rising edge type trigger, we can see two distinctly different rising edge
speeds on this waveform. Also notice that the slower transitioning edge appears
dimmer. This is because the slower transitioning edges occur less often than the
faster transitioning edges. Perhaps these slower transitioning edges are in
violation of meeting minimum required specifications. Let’s first set up the scope
to measure the rise times of this signal, and then we will set up the scope to
uniquely trigger on the violation edges.
9
Press
[Meas]
; then tap the
Type
softkey.
10
Select the
Rise Time
measurement; Once on, tap
Statistics
and then
Display On
.
Figure 31
Scope display reveals two different rising edge speeds when triggering on any
rising edge.