106
4000 X-Series Oscilloscopes Advanced Training Guide
4
Serial Bus Decoding & Triggering, Search & Navigation, and Segmented Acquisition Labs
an automatic search to find all occurrences of Error Frames and “form” errors
(non-flagged error conditions). We will then automatically navigate to each of
these error conditions.
23
Push the horizontal position/delay knob to re-position the trigger point back to
center-screen.
24
Set the scope’s timebase to
100.0 ms/div
.
25
Press the
[Search]
front panel key.
26
Tap the
Search Edge
softkey; then select
Serial 1 (CAN)
.
27
Tap the
Search for
softkey; then select
All Errors
.
The white triangles near the top of the waveform area mark the time location of
each “found” occurrence of our search operation as shown in
. These
frames are also marked in orange in the first/pointer column of the lister table.
Your scope should have found and marked more than 10 occurrences of this
search operation based on a total acquisition time of 1 second. Note that the
number of “found” events is indicated near the bottom of the display. To
automatically navigate to each CAN error, press the
and
front panel
navigation keys.
Figure 70
Using the scope’s automatic Search & Navigation capability to find all occurrences
of CAN errors over a 1 second time span.
Содержание 4000 X-Series
Страница 12: ...12 4000 X Series Oscilloscopes Advanced Training Guide 1 Getting Started...
Страница 156: ...156 4000 X Series Oscilloscopes Advanced Training Guide A Oscilloscope Block Diagram and Theory of Operation...
Страница 166: ...166 4000 X Series Oscilloscopes Advanced Training Guide B Oscilloscope Bandwidth Tutorial...
Страница 168: ...168 4000 X Series Oscilloscopes Advanced Training Guide C Related Keysight Literature...