Oscilloscope Familiarization Labs
2
4000 X-Series Oscilloscopes Advanced Training Guide
41
Because the InfiniiVision oscilloscope’s mask testing capability is hardware-based,
it can test over 200,000 waveforms/sec and provide detailed pass/fail statistics
including failure rate in terms of both percent and a Sigma quality factor as shown
in
. A Sigma quality factor of 6
σ
relates to approximately three or fewer
defects per million tested. Because this particular glitch occurs 1/1,000,000, we
have a Sigma quality factor that exceeds 6
σ
. Let’s now set up the oscilloscope to
stop on the first detected failure.
13
Press
[Run/Stop]
to stop acquisitions.
14
Press the (Back) key.
15
Press
Statistics
.
16
Press
Reset Statistics
to clear the statistics from the previous test.
17
Press the (Back) key.
18
Press
Setup
.
19
Press
On Error – Stop
two times.
20
Press
[Run/Stop]
to begin a new “Stop-on-Error” test.
Figure 20
Mask Testing provides us within detailed statistics about the rate of occurrence of
the infrequent glitch.