Serial Bus Decoding & Triggering, Search & Navigation, and Segmented Acquisition Labs
4
4000 X-Series Oscilloscopes Advanced Training Guide
91
data, tap the
Zoom to Selection
softkey. Let’s now perform an automatic search to
find every occurrence of MOSI data = 03
HEX
. We will then automatically navigate
to each of these occurrences.
35
Push the horizontal position/delay knob to re-position the trigger point back to
center-screen.
36
Set the scope’s timebase to
20.00 ms/div
.
37
Press the
[Search]
front panel key.
38
Tap the
Search Edge
softkey; then select
Serial 1 (SPI)
.
39
Tap the
Bits
softkey.
40
Tap the
Select Digit
softkey; then select the upper nibble (Digit 0).
41
Tap the
Hex
softkey; then turn the Entry knob to select
0
.
42
Tap the
Select Digit
softkey; then turn the Entry knob to select lower nibble (Digit
1).
43
Tap the
Hex
softkey; then turn the Entry knob to select
3
.
The white triangles near the top of the waveform area mark the time location of
each “found” occurrence of our search operation as shown in
. These
frames are also marked in orange in the first/pointer column of the lister table.
Figure 58
Automatic Search & Navigation on SPI traffic.
Содержание 4000 X-Series
Страница 12: ...12 4000 X Series Oscilloscopes Advanced Training Guide 1 Getting Started...
Страница 156: ...156 4000 X Series Oscilloscopes Advanced Training Guide A Oscilloscope Block Diagram and Theory of Operation...
Страница 166: ...166 4000 X Series Oscilloscopes Advanced Training Guide B Oscilloscope Bandwidth Tutorial...
Страница 168: ...168 4000 X Series Oscilloscopes Advanced Training Guide C Related Keysight Literature...