120
4000 X-Series Oscilloscopes Advanced Training Guide
4
Serial Bus Decoding & Triggering, Search & Navigation, and Segmented Acquisition Labs
19
Tap the
Bus Config
softkey; then verify that
Word Size
is set to “8”,
Receiver
is set
to “8”,
Alignment
is set to “Standard I
2
S”,
WS Low
is set to “Left channel”,
SCLK
Slope
is set “rising edge”.
20
Use your finger to drag the data signals to match the order on the screen's help
image by dragging the digital channel identifiers on the far left of the display.
Your scope should now be decoding the left-channel and right-channel audio data
similar to
. However, the scope is not yet triggering on these signals. The
scope should be randomly auto triggering because the default trigger condition is
to trigger on a rising edge of channel-1. But there are no signals present on the
channel-1 input. Let’s now set up the scope to trigger on I
2
S right-channel audio
data as the value of the data increases from
≤
-10 to
≥
+10 (signed decimal
format).
21
Press the
[Trigger]
front panel key; then select
Serial 1 (I
2
S)
using the
Trigger Type
menu.
22
Tap the
Audio Left
softkey; then change to
Right
.
23
Tap the
Trigger =
softkey; then set it to
Increasing
value.
24
Press the
[Mode/Coupling]
key in the Trigger section of the front panel; then
select the
Normal
trigger mode.
Figure 80
Decoding I
2
S signals.
Содержание 4000 X-Series
Страница 12: ...12 4000 X Series Oscilloscopes Advanced Training Guide 1 Getting Started...
Страница 156: ...156 4000 X Series Oscilloscopes Advanced Training Guide A Oscilloscope Block Diagram and Theory of Operation...
Страница 166: ...166 4000 X Series Oscilloscopes Advanced Training Guide B Oscilloscope Bandwidth Tutorial...
Страница 168: ...168 4000 X Series Oscilloscopes Advanced Training Guide C Related Keysight Literature...