118
4000 X-Series Oscilloscopes Advanced Training Guide
4
Serial Bus Decoding & Triggering, Search & Navigation, and Segmented Acquisition Labs
Lab #20: Decoding, Triggering, and Searching on I
2
S Serial Bus
Signals
In this lab you will learn how to set up the scope to decode and trigger on the I
2
S
(audio) serial bus traffic. In addition, you will learn how to use the scope’s
automatic I
2
S Search & Navigation capability, as well as Segmented Memory
acquisition. This lab does not provide a tutorial on the I
2
S protocol and signaling.
To complete this lab, your scope must be licensed with the I
2
S trigger and decode
option (Option AUDIO), as well as the MSO option. You can verify the installed
options on your oscilloscope at
[Help] > About Oscilloscope
.
1
Press
[Default Setup]
on the scope’s front panel.
2
Press
[Help]
; then tap the
Training Signals
softkey.
3
From the
Training Signals
menu, select and enable the
I
2
S
signal.
4
Press the
[1]
front panel key two times to turn off channel-1.
5
Press the
[Digital]
front panel key to turn on the digital channels of acquisition.
6
Tap the
Turn off D7-D0
softkey to turn off the lower 8 digital channels.
7
Tap the
Turn off D15-D9
softkey to turn off the upper 8 digital channels.
8
Tap the
Channel
softkey.
9
Double-tap
D9
to turn on D9.
10
Repeat Step #9 to also turn on D8 and D7.
11
Set the scope’s timebase to
500.0 µs/div
.
Содержание 4000 X-Series
Страница 12: ...12 4000 X Series Oscilloscopes Advanced Training Guide 1 Getting Started...
Страница 156: ...156 4000 X Series Oscilloscopes Advanced Training Guide A Oscilloscope Block Diagram and Theory of Operation...
Страница 166: ...166 4000 X Series Oscilloscopes Advanced Training Guide B Oscilloscope Bandwidth Tutorial...
Страница 168: ...168 4000 X Series Oscilloscopes Advanced Training Guide C Related Keysight Literature...