Oscilloscope Familiarization Labs
2
4000 X-Series Oscilloscopes Advanced Training Guide
19
and 80% threshold levels, or perhaps relative to absolute voltage levels, such as
from/to ± 1.0 V. Let’s now set up our scope to measure just the rise time of this
pulse relative to the 20% and 80% threshold levels.
20
Press the
Clear Meas
softkey; then press the
Clear All
softkey.
21
Press the
Settings
softkey; then press the
Thresholds
softkey.
22
Press the
Lower
softkey; set the value to
20%
.
23
Press the
Upper
softkey; set the value to
80%
.
24
To return to the previous menu, press the (Back) front panel key (just above
the power switch).
25
Press the (Back) key again because we descended two levels into this menu.
26
Press the
Rise
measurement on the right of the display; select
Track with Cursors
.
Using these user-defined measurement threshold levels (20% and 80%), our rise
time measurement should be faster because we are now measuring across a
shorter segment of the waveform as shown in
. The measurement should
now read approximately 30 ns. When we used the scope’s default 10%/90%
threshold levels, the measurement should have read approximately 40 ns.
Let’s now make one more measurement before completing this lab. But this time
let’s perform a more comprehensive set of measurements on this waveform.
Figure 6
Performing a rise time measurement relative to 20% and 80% threshold levels.
Back
Back
Содержание 4000 X-Series
Страница 12: ...12 4000 X Series Oscilloscopes Advanced Training Guide 1 Getting Started...
Страница 156: ...156 4000 X Series Oscilloscopes Advanced Training Guide A Oscilloscope Block Diagram and Theory of Operation...
Страница 166: ...166 4000 X Series Oscilloscopes Advanced Training Guide B Oscilloscope Bandwidth Tutorial...
Страница 168: ...168 4000 X Series Oscilloscopes Advanced Training Guide C Related Keysight Literature...