Advanced Triggering, Search & Navigation, and Segmented Acquisition Labs
3
4000 X-Series Oscilloscopes Advanced Training Guide
59
At this timebase setting (200 µs/div), the scope captured over a thousand edges of
this signal. And using the search conditions we entered (rising edges that exceed
100 ns), the scope should have just found approximately 10 violation edges as
shown in
. The white triangle marks at the top of the display indicate the
location of each “found” violation. Let’s now “navigate” to each of these detected
violations.
24
Press the
(zoom) front panel key.
25
Press the
and
navigation keys to navigate to each discovered edge
violation.
Let’s now use the scope’s Segmented Memory mode of acquisition to capture 500
consecutive occurrences of this edge transition violation. But to complete the rest
of this lab your scope must be licensed with the Segmented Memory option.
Press the (zoom) front key to turn off the scope’s horizontal zoom mode.
26
Push the horizontal position/delay knob to re-position the trigger point back to
center-screen.
27
Set the scope’s timebase to
100.0 ns/div
.
28
Press the
[Run/Stop]
front panel to begin repetitive acquisition again.
29
Press the
[Acquire]
front panel key.
Figure 34
The scope automatically finds several rise time violation edges.
Содержание 4000 X-Series
Страница 12: ...12 4000 X Series Oscilloscopes Advanced Training Guide 1 Getting Started...
Страница 156: ...156 4000 X Series Oscilloscopes Advanced Training Guide A Oscilloscope Block Diagram and Theory of Operation...
Страница 166: ...166 4000 X Series Oscilloscopes Advanced Training Guide B Oscilloscope Bandwidth Tutorial...
Страница 168: ...168 4000 X Series Oscilloscopes Advanced Training Guide C Related Keysight Literature...