Serial Bus Decoding & Triggering, Search & Navigation, and Segmented Acquisition Labs
4
4000 X-Series Oscilloscopes Advanced Training Guide
147
12
Set the scope timebase to
200.0 µs/div
.
13
Press the
[Search]
feature in the Horizontal area of the front panel.
14
Tap the
Search Edge
softkey and change the mode to
Serial 1: MIL-STD-1553
.
15
Open the Lister by tapping the
double down arrows
to the upper right of the
waveform display.
White arrows should appear on the top of the waveform, below the Lister window.
These represent the locations of all hits from the search parameters, namely all
command and segment status word starts. There should be 10-11 on the display.
They are also marked in the Lister window with orange tabs on the left column.
16
Tap
Search for C/S Start
and change it to
Sync Error
.
17
Press
[Run/Stop]
to halt all acquisitions.
18
Use the
and
front panel navigation keys to automatically scroll through
all search results. The lower screen will automatically zoom into the sync error
in the search list, and point to it in the lister window. See
Figure 100
Triggering on a unique sync error.