28
4000 X-Series Oscilloscopes Advanced Training Guide
2
Oscilloscope Familiarization Labs
23
Press the
[Cursors]
front panel key (near the Cursors knob).
24
Press the
Source
softkey and change the
Source
from channel-1 to
Math: f(t)
25
Push the Cursors knob and select the
X1
cursor.
26
Move the X1 cursor so that it is on top of the highest frequency peak (near the
left side of the display).
27
On the right of the screen, select the
X2
cursor.
28
Move the X2 cursor until it is on top of the 2nd highest frequency peak.
The frequencies at the X1 and X2 cursor locations are displayed near the bottom of
the scope’s display.
Let’s now verify the fundamental frequency of this signal using the scope’s
“counter” measurement.
29
Press the
[Meas]
front panel key.
30
Press the
Type
softkey and select
Counter
.
The “counter” measurement measures the average frequency over a fixed gate
time to provide a very high resolution (5 digits) measure of frequency, similar to a
standalone counter. But the measurement requires that the input signal be
repetitive.
The “frequency” measurement measures the period of just one cycle of the signal,
and then computes the frequency by taking the reciprocal.
Содержание 4000 X-Series
Страница 12: ...12 4000 X Series Oscilloscopes Advanced Training Guide 1 Getting Started...
Страница 156: ...156 4000 X Series Oscilloscopes Advanced Training Guide A Oscilloscope Block Diagram and Theory of Operation...
Страница 166: ...166 4000 X Series Oscilloscopes Advanced Training Guide B Oscilloscope Bandwidth Tutorial...
Страница 168: ...168 4000 X Series Oscilloscopes Advanced Training Guide C Related Keysight Literature...