Serial Bus Decoding & Triggering, Search & Navigation, and Segmented Acquisition Labs
4
4000 X-Series Oscilloscopes Advanced Training Guide
107
Let’s now use the scope’s Segmented Memory mode of acquisition to capture 100
consecutive occurrences of CAN error conditions. But to complete the rest of this
lab your scope must be licensed with the Segmented Memory option.
28
Push the horizontal position/delay knob to re-position the trigger point back to
center-screen.
29
Set the scope’s timebase to
200.0 µs/div
.
30
Press the
[Run/Stop]
front panel to begin repetitive acquisition again.
31
Press the
[Trigger]
front panel key.
32
Tap the
Trigger: ID & ~RTR
softkey; then select
All Errors
.
33
Press the
[Mode/Coupling]
key in the Trigger section of the front panel; then
select the
Normal
trigger mode.
34
Press the
[Acquire]
front panel key.
35
Tap the
Segmented
softkey; then enter
100
as the number of segments to
capture.
36
Tap the
Segmented
softkey to begin a Segmented Memory acquisition.
37
Tap the
Current Segment
softkey; then turn the Entry knob to review all 100
captured segments.
38
Set the
Current Segment
to
100
and note the time-tag of the last captured
segment. This can be done with the Entry knob, or double-tapping
Current Seg
and tapping
Max
on the keypad.
Содержание 4000 X-Series
Страница 12: ...12 4000 X Series Oscilloscopes Advanced Training Guide 1 Getting Started...
Страница 156: ...156 4000 X Series Oscilloscopes Advanced Training Guide A Oscilloscope Block Diagram and Theory of Operation...
Страница 166: ...166 4000 X Series Oscilloscopes Advanced Training Guide B Oscilloscope Bandwidth Tutorial...
Страница 168: ...168 4000 X Series Oscilloscopes Advanced Training Guide C Related Keysight Literature...