Serial Bus Decoding & Triggering, Search & Navigation, and Segmented Acquisition Labs
4
4000 X-Series Oscilloscopes Advanced Training Guide
83
21
Set the scope’s timebase to
200.0 µs/div
.
You should now see stable waveform traces on your scope’s display similar to
2
C frame that is near the center of the scope’s display, which is the
frame that the scope is triggering on, should be decoded as:
< 29RA FFA 80~A > or < 29RA FFA C0~A >
This particular frame should be interpreted as a Read (R) operation from address
29 with an acknowledge (A), followed by a data byte equal to FF with an
acknowledge (A), following by a data byte equal to either 80 or C0 without an
acknowledge (~A). Let’s now capture a very long stream of I
2
C data and then
manually search through the captured and decoded record.
22
Set the scope’s timebase to
20.00 ms/div
.
23
Press the
[Serial]
front panel key; then tap the
Lister
softkey.
24
Press
[Run/Stop]
to stop repetitive acquisitions.
25
Tap the
Scroll Lister
softkey; then turn the Entry knob to manually scroll through
the lister table. You can also use the scroll bar to the right of the lister.
As you scroll through the data, note that waveforms “track”. This means that the
frame that the arrow points to in the lister table corresponds to the waveforms that
are positioned at center-screen. If you want to zoom in on a particular I
2
C frame,
Figure 52
Triggering on I
2
C traffic.
Содержание 4000 X-Series
Страница 12: ...12 4000 X Series Oscilloscopes Advanced Training Guide 1 Getting Started...
Страница 156: ...156 4000 X Series Oscilloscopes Advanced Training Guide A Oscilloscope Block Diagram and Theory of Operation...
Страница 166: ...166 4000 X Series Oscilloscopes Advanced Training Guide B Oscilloscope Bandwidth Tutorial...
Страница 168: ...168 4000 X Series Oscilloscopes Advanced Training Guide C Related Keysight Literature...