either push the Entry knob, or tap the
Zoom to Selection
softkey. Let’s now perform
an automatic search to find every occurrence of a “Missing Acknowledge”. We will
then automatically navigate to each of these occurrences.
26
Push the horizontal position/delay knob to re-position the trigger point back to
center-screen.
27
Set the scope’s timebase to
20.00 ms/div
.
28
Press the
[Search]
front panel key.
29
Tap the
Search Edge
softkey; then select
Serial 1 (I
2
C)
.
30
Tap the
Search for
softkey; then select
Missing Acknowledge
as our search criteria.
The white triangles near the top of the waveform area mark the time location of
each “found” occurrence of our search operation as shown in
. These
frames are also marked in orange in the first/pointer column of the lister table.
Your scope should have found and marked approximately 25 occurrences of this
search operation based on a total acquisition time of 200 ms. Note that the
number of “found” events is indicated near the bottom of the display. To
automatically navigate to each frame with a “Missing Acknowledge”, press the
and
front panel navigation keys near the timebase controls.
Figure 53
Automatic Search & Navigation.
Содержание 4000 X-Series
Страница 12: ...12 4000 X Series Oscilloscopes Advanced Training Guide 1 Getting Started...
Страница 156: ...156 4000 X Series Oscilloscopes Advanced Training Guide A Oscilloscope Block Diagram and Theory of Operation...
Страница 166: ...166 4000 X Series Oscilloscopes Advanced Training Guide B Oscilloscope Bandwidth Tutorial...
Страница 168: ...168 4000 X Series Oscilloscopes Advanced Training Guide C Related Keysight Literature...