4000 X-Series Oscilloscopes Advanced Training Guide
3
Advanced Triggering, Search & Navigation, and Segmented Acquisition Labs
Lab #7: Triggering on a Digital Burst using Trigger Holdoff
Signals in the real world of electronics are rarely as simple as repetitive sine waves
and square waves. Establishing unique trigger points on more complex signals
sometimes requires using trigger “hold-off”. In this lab you will learn how to use
the scope’s trigger hold-off capability in order to trigger on a burst of digital
pulses.
1
Connect the channel-1 probe to the Demo 1 terminal and ground.
2
Press
[Default Setup]
on the scope’s front panel.
3
Press
[Help]
; then tap the
Training Signals
softkey.
4
Tap the
Digital Burst
signal from the
Training Signals
menu.
5
Set channel-1’s V/div setting to
1.00 V/div
.
6
Set channel-1’s offset/position to approximately
+1.7 V
in order to center this
waveform on the scope’s display.
7
Push the Trigger Level knob to automatically set the trigger level to
approximately
50%
.
8
Set the scope’s timebase to
20.00 µs/div
.
Figure 22
Attempting to view a burst of pulses while using the scope’s default trigger setup
conditions.
Содержание 4000 X-Series
Страница 12: ...12 4000 X Series Oscilloscopes Advanced Training Guide 1 Getting Started...
Страница 156: ...156 4000 X Series Oscilloscopes Advanced Training Guide A Oscilloscope Block Diagram and Theory of Operation...
Страница 166: ...166 4000 X Series Oscilloscopes Advanced Training Guide B Oscilloscope Bandwidth Tutorial...
Страница 168: ...168 4000 X Series Oscilloscopes Advanced Training Guide C Related Keysight Literature...