96
4000 X-Series Oscilloscopes Advanced Training Guide
4
Serial Bus Decoding & Triggering, Search & Navigation, and Segmented Acquisition Labs
19
While this menu is active, also verity that the # of Bits is set to
8
, the Baud Rate
is set to
19.2 kb/s
, Polarity is set to
Idle Low
, and Bit Order is set to
LSB
first.
20
Press the
[Label]
front panel key (between channel-1 and channel-2 controls) to
turn on the scope’s default labels.
You should now see RS232/UART decoding of these signals on your scope’s
display similar to
, but we still haven’t established stable triggering. The
scope is still triggering on any rising edge of channel-1 (default trigger condition).
Let’s now set up the scope to trigger when the transmit data signal (TX) is equal to
4D
HEX
.
21
Press the
[Trigger]
front panel key; then select
Serial 1 (UART/RS232)
.
22
Tap the
Trigger Setup
softkey; then select
Tx Data
from the
Trigger
menu.
23
Tap the
Data
softkey; then select
0x4D
using the Entry knob.
24
Set the scope’s timebase to
1.000 ms/div
.
Figure 61
Decoding the RS232/UART serial bus.
Содержание 4000 X-Series
Страница 12: ...12 4000 X Series Oscilloscopes Advanced Training Guide 1 Getting Started...
Страница 156: ...156 4000 X Series Oscilloscopes Advanced Training Guide A Oscilloscope Block Diagram and Theory of Operation...
Страница 166: ...166 4000 X Series Oscilloscopes Advanced Training Guide B Oscilloscope Bandwidth Tutorial...
Страница 168: ...168 4000 X Series Oscilloscopes Advanced Training Guide C Related Keysight Literature...