Connections
2-7
WARNING
There is no direct internal connection between earth ground and the
INPUT/OUTPUT LO terminal of the SourceMeter. Therefore, hazardous
voltages (>30V rms) can appear on the LO terminal. Typically, this can
occur when the SourceMeter is operating in any mode where the output
changes rapidly, such as quick, pulsed waveforms that can be generated
using the ZERO, AUTO-OFF output state, or fast pulse sweep operations.
To eliminate the shock hazard, connect the INPUT/OUTPUT LO terminal
to earth ground. You can connect the LO terminal to the chassis ground
screw terminal on the rear panel using the supplied chassis ground plug, or
connect it to a known safety earth ground.
WARNING
For floating source-measure operations, voltage will be present on the
outer shell (IN/OUT LO) of the triax cable(s). If IN/OUT LO is connected
to a noise shield as shown in Figures 2-2 and 2-3, the floating voltage will
also be present on the shield. To prevent contact with the floating voltage,
enclose the test circuit in a metal test fixture that is connected to safety
earth ground (as shown in Figures 2-2 and 2-3), and do not touch the triax
cable(s).
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
INPUT/
OUTPUT
42V
PEAK
250V
PEAK
TRIGGER
LINK
4-WIRE
SENSE
HI
LO
LINE RATING
100-240VAC
50, 60, HZ
100VA MAX
RS232
IEEE-488
(ENTER IEEE ADDRESS
WITH FRONT PANEL MENU)
250V
PEAK
5V
PEAK
5V
PEAK
5V
PK
V,
Ω
,
GUARD
GUARD
SENSE
LINE FUSE
SLOWBLOW
2.5A, 250V
INTERLOCK-
DIGITAL I/O
REMOTE
PreAmp
!
Noise Shield
Triax
Cable
Test
Fixture
HI
Connect to earth safety
ground using #18 AWG
wire or larger
DUT
WARNING Guard is at the same voltage level as
input/output high. Therefore, if a
hazardous voltage (
≥
42V peak) is
present on input/output high, it is also
present on the guard terminals of the
Remote PreAmp and mainframe.
KEITHLEY
6430
REMOTE
PreAmp
SENSE
MAINFRAME
Triax
Cable
LO
Sense
LO
Sense HI
Banana Plug Cable
6430 Mainframe
(rear panel connectors)
IN/OUT
HIGH
Preamp Cable*
*Keithley part number: CA-176-1D.
Figure 2-3
Four-wire sense
connections
Содержание 6430
Страница 26: ......
Страница 32: ......
Страница 78: ...2 14 Connections ...
Страница 98: ...3 20 Basic Source Measure Operation ...
Страница 138: ...5 30 Source Measure Concepts ...
Страница 156: ...6 18 Range Digits Speed and Filters ...
Страница 168: ...7 12 Relative and Math ...
Страница 176: ...8 8 Data Store ...
Страница 202: ...9 26 Sweep Operation ...
Страница 248: ...11 22 Limit Testing ...
Страница 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Страница 418: ...17 108 SCPI Command Reference ...
Страница 450: ...18 32 Performance Verification ...
Страница 477: ...A Specifications ...
Страница 489: ...B StatusandErrorMessages ...
Страница 498: ...B 10 Status and Error Messages ...
Страница 499: ...C DataFlow ...
Страница 503: ...D IEEE 488BusOverview ...
Страница 518: ...D 16 IEEE 488 Bus Overview ...
Страница 519: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 523: ...F MeasurementConsiderations ...
Страница 539: ...G GPIB488 1Protocol ...
Страница 557: ......