SCPI Command Reference
17-61
Repeat
fi
lter commands
:REPeat:COUNt <n>
[:SENSe[1]]:AVERage:REPeat:COUNt <n>
Set repeat
fi
lter count
Parameters
<n> =
1 to 100
Specify repeat
fi
lter count
DEFault
1
MINimum
1
MAXimum
100
Query
:COUNt?
Query
fi
lter count
:COUNt? DEFault
Query the *RST default
fi
lter count
:COUNt? MINimum
Query the lowest allowable
fi
lter count
:COUNt? MAXimum
Query the largest allowable
fi
lter count
Description
This command is used to specify the repeat
fi
lter count. In general, the
fi
lter
count is the number of readings that are acquired and stored in the
fi
lter
buffer for the averaging calculation. Each aquired group of readings yields a
single
fi
ltered reading. The larger the
fi
lter count, the more
fi
ltering that is
performed.
:REPeat[:STATe] <b>
[:SENSe[1]]:AVERage:REPeat[:STATe] <b>
Enable/disable repeat
fi
lter
Parameters
<b> =
0 or OFF
Disable repeat
fi
lter
1 or ON
Enable repeat
fi
lter
Query
[:STATe]?
Query state of repeat
fi
lter
Description
This command is used to enable or disable the repeat
fi
lter. When enabled,
voltage, current, and resistance readings are
fi
ltered according to how the
repeat
fi
lter is con
fi
gured. When disabled, the repeat
fi
lter stage is bypassed.
Median
fi
lter commands
:MEDian:RANK <NRf>
[:SENSe[1]]:MEDian:RANK <NRf>
Set median
fi
lter rank
Parameters
<NRf> =
0 to 5
Specify rank value for median
fi
lter
RANK?
Query median
fi
lter rank
Description
The median
fi
lter is used to pass the
“
middle-most
”
reading from a group of
readings that are arranged according to size. This command is specify the
rank, which determines the number of reading samples for the
fi
lter process.
The number of reading samples are determined as follows:
Sample readings = 2n + 1
Where; n is the selected rank (0 to 5).
Содержание 6430
Страница 26: ......
Страница 32: ......
Страница 78: ...2 14 Connections ...
Страница 98: ...3 20 Basic Source Measure Operation ...
Страница 138: ...5 30 Source Measure Concepts ...
Страница 156: ...6 18 Range Digits Speed and Filters ...
Страница 168: ...7 12 Relative and Math ...
Страница 176: ...8 8 Data Store ...
Страница 202: ...9 26 Sweep Operation ...
Страница 248: ...11 22 Limit Testing ...
Страница 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Страница 418: ...17 108 SCPI Command Reference ...
Страница 450: ...18 32 Performance Verification ...
Страница 477: ...A Specifications ...
Страница 489: ...B StatusandErrorMessages ...
Страница 498: ...B 10 Status and Error Messages ...
Страница 499: ...C DataFlow ...
Страница 503: ...D IEEE 488BusOverview ...
Страница 518: ...D 16 IEEE 488 Bus Overview ...
Страница 519: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 523: ...F MeasurementConsiderations ...
Страница 539: ...G GPIB488 1Protocol ...
Страница 557: ......