SOURCE SPECIFICATIONS
1
Voltage Programming Accuracy (4-wire sense)
2
Accuracy (1 Year)
Noise
Programming
23°C ±5°C
(peak-peak)
Range
Resolution
±% rdg. + volts
0.1Hz – 10Hz
200.000 mV
5 µV
0.02% + 600 µV
5 µV
2.00000
V
50 µV
0.02% + 600 µV
50 µV
20.0000
V
500 µV
0.02% + 2.4 mV
500 µV
200.000
V
5 mV
0.02% +
24 mV
5 mV
TEMPERATURE COEFFICIENT (0
°–
18
°
C & 28
°–
40
°
C):
±(0.15
×
accuracy specification)/°C.
MAX. OUTPUT POWER:
2.2W (four quadrant source or sink operation).
SOURCE/SINK LIMITS:
±21V @ ±105mA, ±210V @ ±10.5mA.
VOLTAGE REGULATION: Line:
0.01% of range.
Load:
0.01% of range + 100µV.
NOISE 10Hz
–
1MHz (p-p):
10mV.
OVER VOLTAGE PROTECTION:
User selectable values, 5% tolerance. Factory default = None.
CURRENT LIMIT:
Bipolar current limit (compliance) set with single value. Min. 0.1% of range.
Current Programming Accuracy (with remote preamp)
Accuracy (1 Year)
1
Noise
Programming
23
°
C ±5
°
C
(peak-peak)
Range
Resolution
±% rdg. +amps
0.1Hz
–
10Hz
1.00000 pA
50 aA
1.0
% +
10 fA
5 fA
10.0000 pA
500 aA
0.50 % +
30 fA
10 fA
100.000 pA
5 fA
0.15 % +
40 fA
20 fA
1.00000 nA
50 fA
0.050 % + 200 f A
50 fA
10.0000 nA
500 fA
0.050 % +
2 pA
500 fA
100.000 nA
5 pA
0.050 % +
20 pA
3 pA
1.00000 µA
50 pA
0.050 % + 300 pA
20 pA
10.0000 µA
500 pA
0.050 % +
2 nA
200 pA
100.000 µA
5 nA
0.031 % + 20 nA
500 pA
1.00000 mA
50 nA
0.034 % + 200 nA
5 nA
10.0000 mA
500 nA
0.045 % +
2 µA
50 nA
100.000 mA
5 µA
0.066 % +
20 µA
500 nA
Current Programming Accuracy (without remote preamp)
Accuracy (1 Year)
1
Noise
Programming
23
°
C ±5
°
C
(peak-peak)
Range
Resolution
±% rdg. +amps
0.1Hz
–
10Hz
1.00000 µA
50 pA
0.035 % + 600 pA
20 pA
10.0000 µA
500 pA
0.033 % +
2 nA
200 pA
100.000 µA
5 nA
0.031 % + 20 nA
500 pA
1.00000 mA
50 nA
0.034 % + 200 nA
5 nA
10.0000 mA
500 nA
0.045 % +
2 µA
50 nA
100.000 mA
5 µA
0.066 % +
20 µA
50 nA
TEMPERATURE COEFFICIENT (0
°–
18
°
C & 28
°–
40
°
C):
±(0.15
×
accuracy specification)/
°
C.
MAX. OUTPUT POWER:
2.2W (four quadrant source or sink operation).
SOURCE/SINK LIMITS:
±10.5mA @ 210V, ±105mA @ 21V.
CURRENT REGULATION: Line:
0.01% of range.
Load:
0.01% of range + 1fA.
VOLTAGE LIMIT:
Bipolar voltage limit (compliance) set with single value. Min. 0.1% of range.
1
For sink mode, 1pA to 100mA range, accuracy is ±(0.15% + offset*4).
2
Voltage source accuracies are not affected by the remote preamp.
Содержание 6430
Страница 26: ......
Страница 32: ......
Страница 78: ...2 14 Connections ...
Страница 98: ...3 20 Basic Source Measure Operation ...
Страница 138: ...5 30 Source Measure Concepts ...
Страница 156: ...6 18 Range Digits Speed and Filters ...
Страница 168: ...7 12 Relative and Math ...
Страница 176: ...8 8 Data Store ...
Страница 202: ...9 26 Sweep Operation ...
Страница 248: ...11 22 Limit Testing ...
Страница 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Страница 418: ...17 108 SCPI Command Reference ...
Страница 450: ...18 32 Performance Verification ...
Страница 477: ...A Specifications ...
Страница 489: ...B StatusandErrorMessages ...
Страница 498: ...B 10 Status and Error Messages ...
Страница 499: ...C DataFlow ...
Страница 503: ...D IEEE 488BusOverview ...
Страница 518: ...D 16 IEEE 488 Bus Overview ...
Страница 519: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 523: ...F MeasurementConsiderations ...
Страница 539: ...G GPIB488 1Protocol ...
Страница 557: ......