SCPI Command Reference
17-67
:AUTO <b>
:SOURce[1]:CURRent:RANGe:AUTO <b>
Select auto range for I-Source
:SOURce[1]:VOLTage:RANGe:AUTO <b>
Select auto range for V-Source
Parameters
<b> =
0 or OFF
Disable auto range
1 or ON
Enable auto range
Query
AUTO?
Query state of auto range
Description
This command is used to enable or disable auto range for the speci
fi
ed
source. When enabled, the instrument will automatically select the most
sensitive range for the speci
fi
ed source level. When disabled, the instrument
will use the range that the instrument is currently on.
Auto range will be disabled if a
fi
xed range is selected. See the previous
command.
Both *RST and :SYSTem:PREset enables source auto range. When the
SourceMeter goes into the local state, source auto range disables.
Set amplitude for
fi
xed source
[:IMMediate][:AMPLitude] <n>
:SOURce[1]:CURRent[:LEVel][:IMMediate][:AMPLitude] <n> Set
fi
xed I-Source amplitude immediately
:SOURce[1]:VOLTage[:LEVel][:IMMediate][:AMPLitude] <n> Set
fi
xed V-Source amplitude immediately
Parameters
<n> =
-105e-3 to 105e-3
Set I-Source amplitude (amps)
-210 to 210
Set V-Source amplitude (volts)
DEFault
0A or 0V
MINimum
-105e-3A or -210V
MAXimum
+105e-3A or +210V
Query
:CURRent?
Query programmed amplitude of I-Source
:CURRent? DEFault
Query *RST default amplitude
:CURRent? MINimum
Query lowest allowable amplitude
:CURRent? MAXimum
Query highest allowable amplitude
:VOLTage?
Query programmed amplitude of V-Source
:VOLTage? DEFault
Query *RST default amplitude
:VOLTage? MINimum
Query lowest allowable amplitude
:VOLTage? MAXimum
Query highest allowable amplitude
Description
This command is used to immediately update the amplitude of a
fi
xed
source. This command is not valid if the list or sweep mode is selected.
NOTE
The sourcing :MODE command is used to select a fixed source. See “Select sourcing
mode.”
Содержание 6430
Страница 26: ......
Страница 32: ......
Страница 78: ...2 14 Connections ...
Страница 98: ...3 20 Basic Source Measure Operation ...
Страница 138: ...5 30 Source Measure Concepts ...
Страница 156: ...6 18 Range Digits Speed and Filters ...
Страница 168: ...7 12 Relative and Math ...
Страница 176: ...8 8 Data Store ...
Страница 202: ...9 26 Sweep Operation ...
Страница 248: ...11 22 Limit Testing ...
Страница 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Страница 418: ...17 108 SCPI Command Reference ...
Страница 450: ...18 32 Performance Verification ...
Страница 477: ...A Specifications ...
Страница 489: ...B StatusandErrorMessages ...
Страница 498: ...B 10 Status and Error Messages ...
Страница 499: ...C DataFlow ...
Страница 503: ...D IEEE 488BusOverview ...
Страница 518: ...D 16 IEEE 488 Bus Overview ...
Страница 519: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 523: ...F MeasurementConsiderations ...
Страница 539: ...G GPIB488 1Protocol ...
Страница 557: ......