19-6
Calibration
Mainframe calibration
Use the procedures discussed below to calibration the mainframe without the Remote
PreAmp. See
Remote PreAmp calibration
later in this section for information in calibrating the
Remote PreAmp.
NOTE
The mainframe must be separately calibrated before calibrating the Remote
PreAmp.
Mainframe calibration menu
You can access the calibration menu by pressing the MENU key and then selecting CAL in
the main menu. Mainframe calibration menu selections include:
•
UNLOCK: Enables calibration.
•
EXECUTE: Allows you to execute the calibration procedure for the selected function
and range.
•
VIEW-DATES: Displays calibration and calibration due dates.
•
SAVE: Saves new calibration constants.
•
LOCK: Disables calibration.
•
CHANGE-PASSWORD: Allows you to change the calibration password.
Mainframe calibration procedure
The mainframe calibration procedure described below calibrates all ranges of both the cur-
rent and voltage source and measure functions. (Resistance calibration is not required.) Note
that each range is separately calibrated by repeating the entire procedure for each range.
Step 1: Prepare the Model 6430 for calibration
1.
With the power off, disconnect the Remote PreAmp from the mainframe.
2.
Turn on the Model 6430 and the digital multimeter, and allow them to warm up for at
least one hour before performing calibration.
3.
Press the MENU key, then choose CAL and press ENTER. Select UNLOCK, and then
press ENTER. The instrument will display the following:
PASSWORD:
Use
,
▲
,
▼
,
, ENTER or EXIT.
4.
Use the up and down range keys to select the letter or number, and use the left and right
arrow keys to choose the position. Enter the present password on the display. (Front
panel default: 006430.) Press ENTER to complete the process.
5.
Press EXIT to return to normal display. Instrument operating states will be set as sum-
marized in Table 19-2.
▲
▲
Содержание 6430
Страница 26: ......
Страница 32: ......
Страница 78: ...2 14 Connections ...
Страница 98: ...3 20 Basic Source Measure Operation ...
Страница 138: ...5 30 Source Measure Concepts ...
Страница 156: ...6 18 Range Digits Speed and Filters ...
Страница 168: ...7 12 Relative and Math ...
Страница 176: ...8 8 Data Store ...
Страница 202: ...9 26 Sweep Operation ...
Страница 248: ...11 22 Limit Testing ...
Страница 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Страница 418: ...17 108 SCPI Command Reference ...
Страница 450: ...18 32 Performance Verification ...
Страница 477: ...A Specifications ...
Страница 489: ...B StatusandErrorMessages ...
Страница 498: ...B 10 Status and Error Messages ...
Страница 499: ...C DataFlow ...
Страница 503: ...D IEEE 488BusOverview ...
Страница 518: ...D 16 IEEE 488 Bus Overview ...
Страница 519: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 523: ...F MeasurementConsiderations ...
Страница 539: ...G GPIB488 1Protocol ...
Страница 557: ......