15-6
Common Commands
*TRG programming example
The command sequence in Table 15-4 con
fi
gures the SourceMeter to be controlled by bus
triggers. The last command, which sends a bus trigger, triggers one measurement. Each subse-
quent bus trigger will also trigger a single measurement.
NOTE
With :ARM:SOURce BUS selected, do not send any commands (except *TRG, GET,
DCL, SDC, IFC, and ABORt) while performing source-measure operations. Other-
wise, erratic operation will occur.
*TST?
—
self-test query
Run self test and read result
Use this query command to perform a checksum test on ROM. The command places the
coded result (0 or 1) in the Output Queue. When the SourceMeter is addressed to talk, the
coded result is sent from the Output Queue to the computer.
A returned value of zero (0) indicates that the test passed, and a value of one (1) indicates
that the test failed.
*WAI
—
wait-to-continue
Wait until previous commands are completed
Effectively, the *WAI command is a No-Op (no operation) for the SourceMeter and thus,
does not need to be used.
Two types of device commands exist:
•
Sequential commands
—
A command whose operations are allowed to
fi
nish before the
next command is executed.
•
Overlapped commands
—
A command that allows the execution of subsequent com-
mands while device operations of the Overlapped command are still in progress.
The *WAI command is used to suspend the execution of subsequent commands until the
device operations of all previous Overlapped commands are
fi
nished. The *WAI command is
not needed for Sequential commands.
Table 15-4
*TRG programming example
Command
Description
*RST
:ARM:SOUR BUS
:ARM:COUN INF
:OUTP ON
:INIT
*TRG
Restore GPIB defaults.
Select BUS control source.
Set arm layer count to in
fi
nite.
Turn on output.
Take SourceMeter out of idle.
Trigger one measurement.
Содержание 6430
Страница 26: ......
Страница 32: ......
Страница 78: ...2 14 Connections ...
Страница 98: ...3 20 Basic Source Measure Operation ...
Страница 138: ...5 30 Source Measure Concepts ...
Страница 156: ...6 18 Range Digits Speed and Filters ...
Страница 168: ...7 12 Relative and Math ...
Страница 176: ...8 8 Data Store ...
Страница 202: ...9 26 Sweep Operation ...
Страница 248: ...11 22 Limit Testing ...
Страница 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Страница 418: ...17 108 SCPI Command Reference ...
Страница 450: ...18 32 Performance Verification ...
Страница 477: ...A Specifications ...
Страница 489: ...B StatusandErrorMessages ...
Страница 498: ...B 10 Status and Error Messages ...
Страница 499: ...C DataFlow ...
Страница 503: ...D IEEE 488BusOverview ...
Страница 518: ...D 16 IEEE 488 Bus Overview ...
Страница 519: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 523: ...F MeasurementConsiderations ...
Страница 539: ...G GPIB488 1Protocol ...
Страница 557: ......