Triggering
10-15
•
HALT
— Use to return the SourceMeter to the idle state. HALT does not turn off the
output. The programmed source level will still be available at the OUTPUT terminals.
The following actions will take the SourceMeter out of idle:
- Turn the output off and then on again.
- Re-select the arm or trigger event.
- Exit from the menu structure, and then re-enter it by pressing CONFIG and then
TRIG.
CONFIG
TRIG
TRIG
LAYER
ARM
LAYER
HALT
IMMEDIATE
ARM IN
COUNT
TRIGGER
IN
TRIGGER
OUT
DELAY
COUNT
LINE
EVENTS
TRIGGER
LINK
LINE
IMMEDIATE
GPIB
TIMER
MANUAL
TLINK
⇓
STEST
⇑
STEST
ARM
OUT
EVENTS
⇑⇓
STEST
Figure 10-8
Configure trigger
menu tree
Содержание 6430
Страница 26: ......
Страница 32: ......
Страница 78: ...2 14 Connections ...
Страница 98: ...3 20 Basic Source Measure Operation ...
Страница 138: ...5 30 Source Measure Concepts ...
Страница 156: ...6 18 Range Digits Speed and Filters ...
Страница 168: ...7 12 Relative and Math ...
Страница 176: ...8 8 Data Store ...
Страница 202: ...9 26 Sweep Operation ...
Страница 248: ...11 22 Limit Testing ...
Страница 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Страница 418: ...17 108 SCPI Command Reference ...
Страница 450: ...18 32 Performance Verification ...
Страница 477: ...A Specifications ...
Страница 489: ...B StatusandErrorMessages ...
Страница 498: ...B 10 Status and Error Messages ...
Страница 499: ...C DataFlow ...
Страница 503: ...D IEEE 488BusOverview ...
Страница 518: ...D 16 IEEE 488 Bus Overview ...
Страница 519: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 523: ...F MeasurementConsiderations ...
Страница 539: ...G GPIB488 1Protocol ...
Страница 557: ......