Sweep Operation
9-13
Setting delay
Generally, the time duration spent at each step (or point) of a sweep consists of the source
delay and the time it takes to perform the measurement (NPLC setting).
The source delay is part of the SDM cycle and is used to allow the source to settle before the
measurement is made. See Section 5,
Source-delay-measure cycle
for details.
The total time period of the source delay could include an auto-delay and/or a user pro-
grammed delay. With auto-delay enabled, 1ms of delay is used. The user programmable source
delay adds 0000.0000 to 9999.9990 seconds of delay. See Section 3,
Source delay
to set these
delays.
Additional delay for a sweep is available by using the trigger delay. This user-specified
delay (0000.0000 to 9999.99990 seconds) occurs before each SDM cycle (device action) of the
sweep. Thus, the trigger delay is executed before each new source-point in the sweep. See
Section 10,
Trigger model
and
Configuring triggering
to set trigger delay.
NOTE
For linear staircase, log staircase, and custom sweeps, source delay, trigger delay,
and NPLC settings are global and affect all sweep points simultaneously. For source
memory sweep only, both the source delay and NPLC settings can be set to different
values for each point in the sweep.
Trigger count and sweep points
The trigger count and number of sweep points should be the same or multiples of one
another. For example, with five sweep points and a trigger count of 10, the sweep will run
twice. See Section 10 for details on trigger count.
Performing sweeps
Procedures for the various sweep types are covered below.
NOTE
The following procedure assumes that the SourceMeter is already connected to the
DUT as explained in Section 2.
WARNING
Hazardous voltages (>30V rms) can appear on the selected INPUT/
OUTPUT LO or Remote PreAmp terminals when performing fast pulse
sweep operations. To eliminate this shock hazard, connect the LO terminal
to earth ground. The ground connection can be made at the chassis ground
screw on the rear panel or to a known safety earth ground.
Содержание 6430
Страница 26: ......
Страница 32: ......
Страница 78: ...2 14 Connections ...
Страница 98: ...3 20 Basic Source Measure Operation ...
Страница 138: ...5 30 Source Measure Concepts ...
Страница 156: ...6 18 Range Digits Speed and Filters ...
Страница 168: ...7 12 Relative and Math ...
Страница 176: ...8 8 Data Store ...
Страница 202: ...9 26 Sweep Operation ...
Страница 248: ...11 22 Limit Testing ...
Страница 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Страница 418: ...17 108 SCPI Command Reference ...
Страница 450: ...18 32 Performance Verification ...
Страница 477: ...A Specifications ...
Страница 489: ...B StatusandErrorMessages ...
Страница 498: ...B 10 Status and Error Messages ...
Страница 499: ...C DataFlow ...
Страница 503: ...D IEEE 488BusOverview ...
Страница 518: ...D 16 IEEE 488 Bus Overview ...
Страница 519: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 523: ...F MeasurementConsiderations ...
Страница 539: ...G GPIB488 1Protocol ...
Страница 557: ......