13-16
Remote Operations
Multiple response messages
If you send more than one query command in the same program message (
Multiple com-
mand messages
), the multiple response messages for all the queries are sent to the computer
when the SourceMeter is addressed to talk. The responses are sent in the order the query com-
mands were sent and are separated by semicolons (;). Items within the same query are sepa-
rated by commas (,). The following example shows the response message for a program
message that contains four single item query commands:
0; 1; 1; 0
Response message terminator (RMT)
Each response is terminated with an LF (Line Feed) and EOI (End Or Identify). The follow-
ing example shows how a multiple response message is terminated:
0; 1; 1; 0 <RMT>
Message exchange protocol
Two rules summarize the message exchange protocol:
Rule 1.
You must always tell the SourceMeter what to send to the computer.
The following two steps must always be performed to send information from the instrument
to the computer:
1.
Send the appropriate query command(s) in a program message.
2.
Address the SourceMeter to talk.
Rule 2.
The complete response message must be received by the computer before another
program message can be sent to the SourceMeter.
RS-232 interface operation
NOTE
The programmable aspects of RS-232 operation (baud rate, data bits, parity, and
terminator) are configured from the COMMUNICATION option of the Main Menu.
See Section 1, “Main Menu.”
Sending and receiving data
The RS-232 interface transfers data using 8 data bits, 1 stop bit, and no parity. Make sure the
device you connect to the SourceMeter also uses these settings.
You can break data transmissions by sending a ^C (decimal 3) or ^X (decimal 18) character
string to the instrument, or by sending an RS-232 break condition (holding the transmit line
low for >11 bits). This clears any pending operation, discards any pending output, and returns a
“
DCL.
”
Содержание 6430
Страница 26: ......
Страница 32: ......
Страница 78: ...2 14 Connections ...
Страница 98: ...3 20 Basic Source Measure Operation ...
Страница 138: ...5 30 Source Measure Concepts ...
Страница 156: ...6 18 Range Digits Speed and Filters ...
Страница 168: ...7 12 Relative and Math ...
Страница 176: ...8 8 Data Store ...
Страница 202: ...9 26 Sweep Operation ...
Страница 248: ...11 22 Limit Testing ...
Страница 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Страница 418: ...17 108 SCPI Command Reference ...
Страница 450: ...18 32 Performance Verification ...
Страница 477: ...A Specifications ...
Страница 489: ...B StatusandErrorMessages ...
Страница 498: ...B 10 Status and Error Messages ...
Страница 499: ...C DataFlow ...
Страница 503: ...D IEEE 488BusOverview ...
Страница 518: ...D 16 IEEE 488 Bus Overview ...
Страница 519: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 523: ...F MeasurementConsiderations ...
Страница 539: ...G GPIB488 1Protocol ...
Страница 557: ......