17-56
SCPI Command Reference
NOTE
The function name must be enclosed in double or single quotes (i.e., :func:stat?
“volt”).
Description
This command is used to query the state of the speci
fi
ed measurement func-
tion. A returned response message of
“
0
”
indicates that the speci
fi
ed func-
tion is disabled, while a
“
1
”
indicates that the function is enabled.
:RESistance:MODE <name>
[:SENSe[1]]:RESistance:MODE <name>
Select ohms measurement mode
Parameters
<name> =
MANual
Manual ohms mode
AUTO
Auto ohms mode
Query
:MODE?
Query ohms mode
Description
This command is used to select the ohms measurement mode. With
MANual ohms selected, the user must con
fi
gure the source and measure
aspects of the operation. When the ohms function is selected, the ohms
reading is simply the result of the V/I calculation.
Range changes cannot be made in manual ohms.
With AUTO ohms selected, the SourceMeter will be con
fi
gured to Source I
Measure V when the ohms function is selected. The current source value
and voltage measurement range used depend on the ohms measurement
range that is selected.
See Section 4 for details on manual and auto ohms.
:RESistance:OCOMpensated <b>
[:SENSe[1]]:RESistance:OCOMpensated <b>
Control offset-compensated ohms
Parameters
<b> =
1 or ON
Enable offset compensation
0 or OFF
Disable offset compensation
Query
:OCOMpensated? Query state of offset compensation
Description
This command is used to enable or disable offset-compensated ohms. When
using the auto ohms measurement mode, the current source level is automat-
ically set. When using the manual ohms measurement mode, you must set
the source (V or I) output level.
See Section 4,
Offset-compensated ohms
for details on making offset-
compensated ohms measurements.
NOTE
Offset-compensated ohms will disable when the :MEASure? command (for the resis-
tance function) or the :CONFigure:RESistance command is sent.
Содержание 6430
Страница 26: ......
Страница 32: ......
Страница 78: ...2 14 Connections ...
Страница 98: ...3 20 Basic Source Measure Operation ...
Страница 138: ...5 30 Source Measure Concepts ...
Страница 156: ...6 18 Range Digits Speed and Filters ...
Страница 168: ...7 12 Relative and Math ...
Страница 176: ...8 8 Data Store ...
Страница 202: ...9 26 Sweep Operation ...
Страница 248: ...11 22 Limit Testing ...
Страница 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Страница 418: ...17 108 SCPI Command Reference ...
Страница 450: ...18 32 Performance Verification ...
Страница 477: ...A Specifications ...
Страница 489: ...B StatusandErrorMessages ...
Страница 498: ...B 10 Status and Error Messages ...
Страница 499: ...C DataFlow ...
Страница 503: ...D IEEE 488BusOverview ...
Страница 518: ...D 16 IEEE 488 Bus Overview ...
Страница 519: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 523: ...F MeasurementConsiderations ...
Страница 539: ...G GPIB488 1Protocol ...
Страница 557: ......