Basic Source-Measure Operation
Remote command measure only
Table 3-8 summarizes the basic command sequence for measure only. The steps outlined
correspond to those in the
Front panel measure only
sequence above. These commands set up
the SourceMeter for measure only voltage measurements up to 20V as follows:
•
Measure function: volts
•
Source function: current
•
Source mode: fixed
•
Source range: minimum
•
Source value: 0mA
•
Measure range: 20V
•
Compliance 25V
T
a
b
le 3
-8
Measure only programming example
Step
1
Action
Commands
2,3
Comments
*RST
Restore GPIB defaults.
1
Select measure, source functions.
:SOUR:FUNC CURR
Current source function.
:SOUR:CURR:MODE FIXED
Fixed current source mode.
:SENS:FUNC “VOLT”
Volts measure function.
2
Set source and compliance.
:SOUR:CURR:RANG MIN
Lowest source range.
:SOUR:CURR:LEV 0
0
µ
A source level.
:SENS:VOLT:PROT 25
25V compliance.
3
Select volts measure range.
:SENS:VOLT:RANG 20
20V range.
5
Turn on output.
:OUTP ON
Output on before measuring.
6
Read data.
:READ?
Trigger, acquire reading.
7
Turn off output.
:OUTP OFF
Output off after measuring.
1
Steps correspond to front panel steps listed previously in
Front panel measure only
. DUT should be connected to SourceMeter before
running program.
2
Commands must be sent in order given.
3
Instrument must be addressed to talk after :READ? to acquire data.
Содержание 6430
Страница 26: ......
Страница 32: ......
Страница 78: ...2 14 Connections ...
Страница 98: ...3 20 Basic Source Measure Operation ...
Страница 138: ...5 30 Source Measure Concepts ...
Страница 156: ...6 18 Range Digits Speed and Filters ...
Страница 168: ...7 12 Relative and Math ...
Страница 176: ...8 8 Data Store ...
Страница 202: ...9 26 Sweep Operation ...
Страница 248: ...11 22 Limit Testing ...
Страница 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Страница 418: ...17 108 SCPI Command Reference ...
Страница 450: ...18 32 Performance Verification ...
Страница 477: ...A Specifications ...
Страница 489: ...B StatusandErrorMessages ...
Страница 498: ...B 10 Status and Error Messages ...
Страница 499: ...C DataFlow ...
Страница 503: ...D IEEE 488BusOverview ...
Страница 518: ...D 16 IEEE 488 Bus Overview ...
Страница 519: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 523: ...F MeasurementConsiderations ...
Страница 539: ...G GPIB488 1Protocol ...
Страница 557: ......