SCPI Command Reference
17-75
:POINts <n>
:SOURce[1]:SWEep:POINts <n>
Set source-measure points for sweep
Parameters
<n> =
1 to 2500
Specify number of source-measure points
MINimum
1
MAXimum
2500
DEFault
2500
Query
:POINts?
Query number of sweep points
:POINts? DEFault
Query *RST default number of sweep points
:POINts? MINimum
Query lowest allowable number of sweep
points
:POINts? MAXimum
Query highest allowable number of sweep
points
Description
The :POINts command speci
fi
es the total number of source-measure points
in a sweep. For a linear sweep, the source-measure points are equally spaced
(stepped) between the start level and the stop level. For a log sweep, the
source-measure points are equally spaced on a logarithmic scale. Note that
the start and stop levels are source-measure points.
Step size for a linear sweep can be calculated as follows:
Step Size = (Stop - Start) / (Points - 1)
Step Size = Span / (Points -1)
Step size for a logarithmic sweep can be calculated as follows:
An alternate way to set the source-measure points in a sweep is to specify
the step size using the :STEP command.
Note that the :POINts and :STEP commands are coupled. Changing the
number of source-measure points also changes the step size. Conversely,
changing the step size changes the number of source-measure points.
:DIRection <name>
:SOURce[1]:SWEep:DIRection <name>
Set direction of sweep
Parameters
<name> =
UP
Run sweep from start to stop
DOWn
Run sweep from stop to start
Query
:DIRection?
Query direction of sweep
Description
Normally, a sweep is run from the start level to the stop level. The :STARt
and :STOP, or :CENTer and :SPAN commands are used to set these levels.
This command lets you change the execution direction of the sweep. With
DOWn selected, the sweep will begin at the stop level and end at the start
level. Selecting UP restores sweep operation to the normal start to stop
direction.
Log Step Size
log 10 (Stop)- log10 (Start)
Points - 1
-----------------------------------------------------------------
=
Содержание 6430
Страница 26: ......
Страница 32: ......
Страница 78: ...2 14 Connections ...
Страница 98: ...3 20 Basic Source Measure Operation ...
Страница 138: ...5 30 Source Measure Concepts ...
Страница 156: ...6 18 Range Digits Speed and Filters ...
Страница 168: ...7 12 Relative and Math ...
Страница 176: ...8 8 Data Store ...
Страница 202: ...9 26 Sweep Operation ...
Страница 248: ...11 22 Limit Testing ...
Страница 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Страница 418: ...17 108 SCPI Command Reference ...
Страница 450: ...18 32 Performance Verification ...
Страница 477: ...A Specifications ...
Страница 489: ...B StatusandErrorMessages ...
Страница 498: ...B 10 Status and Error Messages ...
Страница 499: ...C DataFlow ...
Страница 503: ...D IEEE 488BusOverview ...
Страница 518: ...D 16 IEEE 488 Bus Overview ...
Страница 519: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 523: ...F MeasurementConsiderations ...
Страница 539: ...G GPIB488 1Protocol ...
Страница 557: ......