17-36
SCPI Command Reference
Query
:SOURce2?
Query programmed source value
Description
This command is used to de
fi
ne the 3-bit or 4-bit output pattern for the
Digital I/O Port when a test (limit 2, 3, 5-12) for the sorting mode passes.
Note that the output value can be speci
fi
ed in binary, octal, decimal, or
hexadecimal format. Use the table provided in
:SOURce2<NRf> | <NDN>
,
Description
to determine the parameter value for the desired decimal digital
output pattern.
NOTE
16-bit I/O is available with the 2499-DIGIO option. The maximum 16-bit output
value is 65535.
:STATe <b>
:CALCulate2:LIMit[1]:STATe <b>
Control LIMIT 1 test
:CALCulate2:LIMitx:STATe <b>
Control LIMIT x test (x = 2, 3, 5-12)
Parameters
<b> =
1 or ON
Enable speci
fi
ed limit test
0 or OFF
Disable speci
fi
ed limit test
Query
:STATe?
Query state of speci
fi
ed limit test
Description
These commands are used to enable or disable LIMIT 1, LIMIT 2, LIMIT 3,
and LIMIT 5 to LIMIT 12 tests. Any limit test not enabled is simply not
performed.
When a limit test is enabled, the Digital I/O port comes under control of
limit tests. That is, the result of the testing process updates the output pat-
tern on the I/O port.
:FAIL?
:CALCulate2:LIMit[1]:FAIL?
Read LIMIT 1 test result
:CALCulate2:LIMitx:FAIL?
Read LIMIT x test result (x = 2, 3, 5-12)
Description
These commands are used to read the results of LIMIT 1, LIMIT 2, LIMIT
3, and LIMIT 5 to LIMIT 12 tests:
0 = Limit test passed
1 = Limit test failed
The response message (0 or 1) only tells you if a limit test has passed or
failed. For Limit 2, Limit 3, and Limit 5-12, it does not tell you which limit
(upper or lower) has failed. To determine which limit has failed, you will
have to read the Measurement Event Register. See
STATus subsystem
.
Reading the results of a limit test does not clear the fail indication of the
test. A failure can be cleared by using a :CLEar command.
Содержание 6430
Страница 26: ......
Страница 32: ......
Страница 78: ...2 14 Connections ...
Страница 98: ...3 20 Basic Source Measure Operation ...
Страница 138: ...5 30 Source Measure Concepts ...
Страница 156: ...6 18 Range Digits Speed and Filters ...
Страница 168: ...7 12 Relative and Math ...
Страница 176: ...8 8 Data Store ...
Страница 202: ...9 26 Sweep Operation ...
Страница 248: ...11 22 Limit Testing ...
Страница 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Страница 418: ...17 108 SCPI Command Reference ...
Страница 450: ...18 32 Performance Verification ...
Страница 477: ...A Specifications ...
Страница 489: ...B StatusandErrorMessages ...
Страница 498: ...B 10 Status and Error Messages ...
Страница 499: ...C DataFlow ...
Страница 503: ...D IEEE 488BusOverview ...
Страница 518: ...D 16 IEEE 488 Bus Overview ...
Страница 519: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 523: ...F MeasurementConsiderations ...
Страница 539: ...G GPIB488 1Protocol ...
Страница 557: ......