Limit Testing
11-3
Data
fl
ow
All limit tests are part of the CALC2 data block. See Appendix C for an overview on how
limit testing
fi
ts into the overall data
fl
ow through the SourceMeter.
Limit 1 test (compliance)
This hardware (H/W) test checks the compliance state of the SourceMeter. It uses the pro-
grammed compliance as the test limit. At or above the programmed limit, the instrument is in
compliance. Below the limit, the instrument is not in compliance.
For example, assume you want to
“
pass
”
resistors that are below 1k
Ω
. To do this set the
I-Source to output 1mA at a compliance limit of 1V, and con
fi
gure the test to fail on compli-
ance. If, for example, the resistor under test is 750
Ω
, output voltage will be 0.75V
(1mA
×
750
Ω
= 0.75V). Since the output voltage is below the 1V limit, the test passes. If the
resistor is 1k
Ω
(or more), output voltage will be 1V (1mA
×
1k
Ω
= 1V). Since the 1V limit is
reached, which places the SourceMeter in compliance, the test fails.
The Limit 1 test can be used to determine the polarity of a device, such as a diode. By using
this test with a source memory sweep, you can branch to a different setup at a speci
fi
ed mem-
ory location when the device is installed backwards. See the programming example at the end
of this section for details on diode testing.
Limit 2, limit 3, and limit 5-12 tests
These software (S/W) tests are used to determine if a DUT is within speci
fi
ed high and low
limits. Typically, the Limit 2 test is used to test for
“
coarse
”
tolerance limits, and the Limit 3
and Limit 5-12 tests are used for
“fi
ne
”
tolerance limits.
For example, assume you want to sort resistors into three groups: 1%, 5%, and >5% toler-
ance. To do this, con
fi
gure Limit 2 test for 5% HI and LO limits, and Limit 3 test for 1% HI and
LO limits. If Limit 2 fails, the handler places the DUT in the bin labeled >5%. If Limit 2
passes, Limit 3 test is run. If Limit 3 fails, the DUT is placed in the bin labeled 5%. If Limit 3
passes, the handler places the DUT in the bin labeled 1%.
NOTE
Limit 2 can be used with the percent deviation math function. See “Percent devia-
tion” in Section 7 for details.
Содержание 6430
Страница 26: ......
Страница 32: ......
Страница 78: ...2 14 Connections ...
Страница 98: ...3 20 Basic Source Measure Operation ...
Страница 138: ...5 30 Source Measure Concepts ...
Страница 156: ...6 18 Range Digits Speed and Filters ...
Страница 168: ...7 12 Relative and Math ...
Страница 176: ...8 8 Data Store ...
Страница 202: ...9 26 Sweep Operation ...
Страница 248: ...11 22 Limit Testing ...
Страница 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Страница 418: ...17 108 SCPI Command Reference ...
Страница 450: ...18 32 Performance Verification ...
Страница 477: ...A Specifications ...
Страница 489: ...B StatusandErrorMessages ...
Страница 498: ...B 10 Status and Error Messages ...
Страница 499: ...C DataFlow ...
Страница 503: ...D IEEE 488BusOverview ...
Страница 518: ...D 16 IEEE 488 Bus Overview ...
Страница 519: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 523: ...F MeasurementConsiderations ...
Страница 539: ...G GPIB488 1Protocol ...
Страница 557: ......