Digital I/O Port, Interlock, and Output Configuration
12-9
GUARD
With this output-off state, the current source is selected and set to 0A. Voltage compliance is
set to 0.5% full scale of the present voltage range. This output-off state should be used when
performing 6-wire guarded ohms measurements or for any other load that uses an active
source.
NOTE
When changing the output-off state with the output off, the selected output-off state
will be entered immediately.
Output off states and inductive loads
The output off state you select for inductive loads depends on how much energy the inductor
holds. The Normal output off state is
not
recommended as it lowers the compliance setting. The
Zero or the Guard state are better suited, as Zero does
not
change the compliance setting and
the Guard output off state would change the voltage source to a current source with a voltage
compliance. The Guard state is typically used only for guarded ohms measurements.
To protect the unit from inductive energy, the application may require a spark gap across the
INPUT HI and LO terminals. The SourceMeter does not have internal spark gap protection, as
some leakage current (nA) is associated with the protection circuits.
Remote output con
fi
guration
Output con
fi
guration commands
Table 12-2 summarizes output con
fi
guration commands. These commands include those to
enable and disable the interlock as well as commands to control output off states. See Section
17,
OUTPut subsystem
and
SOURce subsystem
for more information.
Table 12-2
Output configuration commands
Command
Description
:OUTPut:INTerlock:STATe <state>
:OUTPut:INTerlock:TRIPped?
:OUTPut:SMODe <name>
:SOURce:CLEar
:SOURce:CLEar:AUTO <state>
:SOURce:CLEar:AUTO:MODE <name>
Enable/disable interlock (state = ON or OFF).
Query interlock tripped state (1 = tripped).
Select output-off mode (state = NORMal, ZERO, or GUARd).
Turn output source off when in idle state.
Enable/disable auto output-off. State = ON (output off after
measurement) or ON (output stays on).
Auto clear mode. Name = ALWays (every reading; default) or
TCOunt (ON when trigger layer entered; OFF when leaving
trigger layer).
Содержание 6430
Страница 26: ......
Страница 32: ......
Страница 78: ...2 14 Connections ...
Страница 98: ...3 20 Basic Source Measure Operation ...
Страница 138: ...5 30 Source Measure Concepts ...
Страница 156: ...6 18 Range Digits Speed and Filters ...
Страница 168: ...7 12 Relative and Math ...
Страница 176: ...8 8 Data Store ...
Страница 202: ...9 26 Sweep Operation ...
Страница 248: ...11 22 Limit Testing ...
Страница 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Страница 418: ...17 108 SCPI Command Reference ...
Страница 450: ...18 32 Performance Verification ...
Страница 477: ...A Specifications ...
Страница 489: ...B StatusandErrorMessages ...
Страница 498: ...B 10 Status and Error Messages ...
Страница 499: ...C DataFlow ...
Страница 503: ...D IEEE 488BusOverview ...
Страница 518: ...D 16 IEEE 488 Bus Overview ...
Страница 519: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 523: ...F MeasurementConsiderations ...
Страница 539: ...G GPIB488 1Protocol ...
Страница 557: ......