SCPI Command Reference
17-33
Con
fi
gure and control limit tests
:COMPliance:FAIL <name>
:CALCulate2:LIMit[1]:COMPliance:FAIL <name>
Parameters
<name> =
IN
Fail Limit 1 test when unit goes into
compliance
OUT
Fail Limit 1 test when unit comes out of
compliance
Query
:FAIL?
Query when Limit 1 test failure occurs
Description
This command is used to specify the condition that will cause Limit 1 test to
fail. With IN speci
fi
ed, the test will fail when the SourceMeter goes into
compliance. With OUT speci
fi
ed, the test will fail when the SourceMeter
comes out of compliance.
[:DATA] <n>
:CALCulate2:LIMitx:LOWer[:DATA] <n>
Specify lower LIMIT x (x = 2, 3, 5-12)
:CALCulate2:LIMitx:UPPer[:DATA] <n>
Specify upper LIMIT x (x = 2, 3, 5-12)
Parameters
<n> =
-9.999999e20 to
Specify limit value
9.999999e20
DEFault
Set speci
fi
ed lower limit to -1
Set speci
fi
ed upper limit to 1
MINimum
Set speci
fi
ed limit to -9.999999e20
MAXimum
Set speci
fi
ed limit to +9.999999e20
Query
:UPPer?
Query speci
fi
ed upper limit
:UPPer? DEFault
Query *RST default upper limit
:UPPer? MINimum
Query lowest allowable upper limit
:UPPer? MAXimum
Query largest allowable upper limit
:LOWer?
Query speci
fi
ed lower limit
:LOWer? DEFault
Query *RST default lower limit
:LOWer? MINimum
Query lowest allowable lower limit
:LOWer? MAXimum
Query largest allowable lower limit
Description
These commands are used to set the upper and lower limits for LIMIT 2,
LIMIT 3, and LIMIT 5 through LIMIT 12 tests. The actual limit depends on
which measurement function is currently selected. For example, a limit
value of 1µ is 1µA for the amps function and 1µV for the v olts function. A
limit value is not range sensitive. A limit of 2 for volts is 2V on all measure-
ment ranges.
Содержание 6430
Страница 26: ......
Страница 32: ......
Страница 78: ...2 14 Connections ...
Страница 98: ...3 20 Basic Source Measure Operation ...
Страница 138: ...5 30 Source Measure Concepts ...
Страница 156: ...6 18 Range Digits Speed and Filters ...
Страница 168: ...7 12 Relative and Math ...
Страница 176: ...8 8 Data Store ...
Страница 202: ...9 26 Sweep Operation ...
Страница 248: ...11 22 Limit Testing ...
Страница 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Страница 418: ...17 108 SCPI Command Reference ...
Страница 450: ...18 32 Performance Verification ...
Страница 477: ...A Specifications ...
Страница 489: ...B StatusandErrorMessages ...
Страница 498: ...B 10 Status and Error Messages ...
Страница 499: ...C DataFlow ...
Страница 503: ...D IEEE 488BusOverview ...
Страница 518: ...D 16 IEEE 488 Bus Overview ...
Страница 519: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 523: ...F MeasurementConsiderations ...
Страница 539: ...G GPIB488 1Protocol ...
Страница 557: ......