SCPI Command Reference
17-79
:POINts <NRf>
:SOURCe:MEMory:POINts <NRf>
Specify number of sweep points to exe-
cute
Parameters
<NRf> = 1 to 100
Number of sweep points
Description
This command is used to specify the number of points for the sweep. For
example, if you saved setups in memory locations 1 through 12 for a sweep,
specify a 12-point sweep using this command.
:STARt <NRf>
:SOURCe:MEMory:STARt <NRf>
Select Source Memory Sweep start
location
Parameters
<NRf> = 1 to 100
Specify memory location
Description
This command is used to set the starting location of a Source Memory
Sweep. For example, for setups saved in memory locations 98 through 5,
specify a starting location of 98.
:RECall <NRf>
:SOURCe:MEMory:RECall <NRf>
Return to speci
fi
ed setup
Parameters
<NRf> = 1 to 100
Specify memory location
Description
This command is used to return the SourceMeter to the setup stored at the
speci
fi
ed memory location.
Set scaling factor
:TRIGgered:SFACtor <n>
:SOURce[1]:CURRent[:LEVel]:TRIGgered:SFACtor <n>
Set current scaling factor
:SOURce[1]:VOLTage[:LEVel]:TRIGgered:SFACtor <n>
Set voltage scaling factor
Parameters
<n> =
-999.9999e+18 to 999.9999e+18
Scaling factor
Query
:SFACTor?
Query scaling factor
Description
:SFAC instructs the SourceMeter to source the scaling factor times the pre-
vious source memory location value. For example, if 10.0V is stored in the
fi
rst source memory (Source I, Measure V Mode), and the unit is in the
Source V, Measure I mode with :SFAC set to 0.1 and enabled, the unit will
output 1.0V for the second source memory location.
NOTE
These commands work only with source memory sweeps.
Содержание 6430
Страница 26: ......
Страница 32: ......
Страница 78: ...2 14 Connections ...
Страница 98: ...3 20 Basic Source Measure Operation ...
Страница 138: ...5 30 Source Measure Concepts ...
Страница 156: ...6 18 Range Digits Speed and Filters ...
Страница 168: ...7 12 Relative and Math ...
Страница 176: ...8 8 Data Store ...
Страница 202: ...9 26 Sweep Operation ...
Страница 248: ...11 22 Limit Testing ...
Страница 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Страница 418: ...17 108 SCPI Command Reference ...
Страница 450: ...18 32 Performance Verification ...
Страница 477: ...A Specifications ...
Страница 489: ...B StatusandErrorMessages ...
Страница 498: ...B 10 Status and Error Messages ...
Страница 499: ...C DataFlow ...
Страница 503: ...D IEEE 488BusOverview ...
Страница 518: ...D 16 IEEE 488 Bus Overview ...
Страница 519: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 523: ...F MeasurementConsiderations ...
Страница 539: ...G GPIB488 1Protocol ...
Страница 557: ......