SCPI Command Reference
17-31
CALCulate2
Con
fi
gure and control limit tests
The following commands are used to con
fi
gure and control the three limit tests for DUT.
When used with a handler to provide binning operations, communication between the
SourceMeter and the handler is provided via the Digital I/O port. Many control aspects of the
digital output lines are performed from the SOURce2 Subsystem. These control aspects include
setting and clearing the digital output lines, and setting pulse width. See
SOURce2
for details.
Select input path
:FEED <name>
:CALCulate2:FEED <name>
Select input path for limit tests
Parameters
<name> = CALCulate[1]
Use result of CALC1
VOLTage
Use measured voltage reading
CURRent
Use measured current reading
RESistance
Use measured resistance reading
Query
:FEED?
Query input path for limit tests
Description
This command is used to select the input path for the limit tests. With
CALCulate[1] selected, the speci
fi
ed limits will be compared to the result of
CALC1. With VOLTage selected, limits will be compared to the voltage
measurement. With CURRent or RESistance selected, limits will be
compared with the respective current or resistance measurement.
Null feed reading
:OFFSet <n>
:CALCulate2:NULL:OFFSet <n>
Specify null offset (REL) for feed
Parameters
<n> = -9.999999e20 to
Specify null offset value
9.999999e20
Query
:OFFSet?
Query null offset value
Description
This command lets you establish a null offset (REL) for the selected feed.
When Null Offset is enabled (see :Ingolstadt), the result is the algebraic dif-
ference between the feed reading and the offset value:
CALC2 reading = feed reading - null offset
Содержание 6430
Страница 26: ......
Страница 32: ......
Страница 78: ...2 14 Connections ...
Страница 98: ...3 20 Basic Source Measure Operation ...
Страница 138: ...5 30 Source Measure Concepts ...
Страница 156: ...6 18 Range Digits Speed and Filters ...
Страница 168: ...7 12 Relative and Math ...
Страница 176: ...8 8 Data Store ...
Страница 202: ...9 26 Sweep Operation ...
Страница 248: ...11 22 Limit Testing ...
Страница 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Страница 418: ...17 108 SCPI Command Reference ...
Страница 450: ...18 32 Performance Verification ...
Страница 477: ...A Specifications ...
Страница 489: ...B StatusandErrorMessages ...
Страница 498: ...B 10 Status and Error Messages ...
Страница 499: ...C DataFlow ...
Страница 503: ...D IEEE 488BusOverview ...
Страница 518: ...D 16 IEEE 488 Bus Overview ...
Страница 519: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 523: ...F MeasurementConsiderations ...
Страница 539: ...G GPIB488 1Protocol ...
Страница 557: ......