Status Structure
Measurement Event Register
The used bits of the Measurement Event Register (shown in Figure 14-6) are described as
follows:
•
Bit B0, Limit 1 Fail (L1)
—
Set bit indicates that the Limit 1 test has failed.
•
Bit B1, Low Limit 2 Fail (LL2)
—
Set bit indicates that the Low Limit 2 test has failed.
•
Bit B2, High Limit 2 Fail (HL2)
—
Set bit indicates that the High Limit 2 test has
failed.
•
Bit B3, Low Limit 3 Fail (LL3)
—
Set bit indicates that the Low Limit 3 test has failed.
•
Bit B4, High Limit 3 Fail (HL3)
—
Set bit indicates that the High Limit 3 test has
failed.
•
Bit B5, Limits Pass (LP)
—
Set bit indicates that all limit tests passed.
•
Bit B6, Reading Available (RAV)
—
Set bit indicates that a reading was taken and
processed.
•
Bit B7, Reading Overflow (ROF)
—
Set bit indicates that the volts or amps reading
exceeds the selected measurement range of the SourceMeter.
•
Bit B8, Buffer Available (BAV)
—
Set bit indicates that there are at least two readings
in the buffer.
•
Bit B9, Buffer Full (BFL)
—
Set bit indicates that the trace buffer is full.
•
Bit B10
—
Not used.
•
Bit B11, Interlock Asserted (Int)
—
Set bit indicates that the interlock line is at digital
low (asserted). The source output can be turned on.
•
Bit B12, Over Temperature (OT)
—
Set bit indicates that an over temperature condi-
tion exists. The source output cannot be turned on.
•
Bit B13, Over Voltage Protection (OVP)
—
Set bit indicates that the source is being
limited at the programmed limit level.
•
Bit B14, Compliance (Comp)
—
Set bit indicates that the source is in compliance.
•
Bit B15
—
Not used.
Содержание 6430
Страница 26: ......
Страница 32: ......
Страница 78: ...2 14 Connections ...
Страница 98: ...3 20 Basic Source Measure Operation ...
Страница 138: ...5 30 Source Measure Concepts ...
Страница 156: ...6 18 Range Digits Speed and Filters ...
Страница 168: ...7 12 Relative and Math ...
Страница 176: ...8 8 Data Store ...
Страница 202: ...9 26 Sweep Operation ...
Страница 248: ...11 22 Limit Testing ...
Страница 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Страница 418: ...17 108 SCPI Command Reference ...
Страница 450: ...18 32 Performance Verification ...
Страница 477: ...A Specifications ...
Страница 489: ...B StatusandErrorMessages ...
Страница 498: ...B 10 Status and Error Messages ...
Страница 499: ...C DataFlow ...
Страница 503: ...D IEEE 488BusOverview ...
Страница 518: ...D 16 IEEE 488 Bus Overview ...
Страница 519: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 523: ...F MeasurementConsiderations ...
Страница 539: ...G GPIB488 1Protocol ...
Страница 557: ......