18-28
Performance Verification
4.
Verify ohms measurement accuracy for each of the resistance values listed in
Table 18-14. For each measurement:
•
Set the resistance calibrator output to the nominal resistance or closest available
value.
NOTE
It may not be possible to set the resistance calibrator to the specified value. Use the
closest possible setting, and modify reading limits accordingly.
•
Select the appropriate ohms measurement range with the RANGE keys.
•
Verify that the Model 6430 resistance reading is within the limits given in the
table.
2G
Ω
-200G
Ω
range accuracy
1.
With the power off, connect the Remote PreAmp IN/OUT HIGH jack directly to the
Model 5156 OUTPUT jack as shown in Figure 18-9. (Use male triax-to-triax adapters.
Also, be sure to remove the link between Model 5156 SHIELD and CHASSIS, and
connect the BNC shorting cap to the resistance cap being used.
)
Table 18-14
Remote PreAmp 20
Ω
-200M
Ω
range measurement accuracy limits
Model 6430 range
Calibrator resistance
1
Model 6430 resistance reading limits
2
(1 year, 18°-28°C)
20
Ω
200
Ω
2k
Ω
20k
Ω
200k
Ω
2M
Ω
20M
Ω
200M
Ω
19
Ω
190
Ω
1.9k
Ω
19k
Ω
190k
Ω
1.9M
Ω
19M
Ω
100M
Ω
18.9784 to 19.0216
Ω
189.824 to 190.176
Ω
1.89845 to 1.90155k
Ω
18.9850 to 19.0150k
Ω
189.814 to 190.186k
Ω
1.89814 to 1.90186M
Ω
18.9829 to 19.0171M
Ω
99.905
to 100.095M
Ω
1
Nominal resistance value.
2
Reading limits based on Model 6430 normal accuracy speci
fi
cations and nominal resistance values. If actual
resistance values differ from nominal values shown, recalculate reading limits using actual calibrator resis-
tance values and Model 6430 normal accuracy speci
fi
cations. See
Verification limits
earlier in this section for
details.
Содержание 6430
Страница 26: ......
Страница 32: ......
Страница 78: ...2 14 Connections ...
Страница 98: ...3 20 Basic Source Measure Operation ...
Страница 138: ...5 30 Source Measure Concepts ...
Страница 156: ...6 18 Range Digits Speed and Filters ...
Страница 168: ...7 12 Relative and Math ...
Страница 176: ...8 8 Data Store ...
Страница 202: ...9 26 Sweep Operation ...
Страница 248: ...11 22 Limit Testing ...
Страница 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Страница 418: ...17 108 SCPI Command Reference ...
Страница 450: ...18 32 Performance Verification ...
Страница 477: ...A Specifications ...
Страница 489: ...B StatusandErrorMessages ...
Страница 498: ...B 10 Status and Error Messages ...
Страница 499: ...C DataFlow ...
Страница 503: ...D IEEE 488BusOverview ...
Страница 518: ...D 16 IEEE 488 Bus Overview ...
Страница 519: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 523: ...F MeasurementConsiderations ...
Страница 539: ...G GPIB488 1Protocol ...
Страница 557: ......