10-10
Triggering
The Trigger Link connections for this test system are shown in Figure 10-6. Trigger Link of
the SourceMeter is connected to Trigger Link (IN or OUT) of the switching mainframe. Note
that with the default trigger settings of the switching mainframe, line #1 is an input, and line #2
is an output.
For this example, the SourceMeter and Switching Mainframe are configured as follows.
SourceMeter setup
Step 1: Restore bench defaults
Press the MENU key, select SAVESETUP, then press the ENTER key. From the
SAVESETUP menu, select GLOBAL, then press ENTER. From the GLOBAL SETUP
MENU, select RESET, then press ENTER. Select BENCH, then press ENTER.
Step 2: Set up trigger parameters
Press the CONFIG key, and then the TRIG key to access the CONFIGURE TRIGGER
menu. Select TRIG-LAYER, and then press ENTER to access the CONFIGURE TRIG-
LAYER menu.
Step 3: Set trigger-in event to TRIGGER-LINK
Select TRIGGER-IN, then press ENTER. Select TRIGGER-LINK, then press ENTER.
Step 4: Set trigger input line to #2
Select #2, then press ENTER three times to return to CONFIGURE TRIG-LAYER menu.
Step 5: Set trigger output line to # 1
Select TRIGGER-OUT, then press ENTER. Select LINE, then press ENTER. Select #1,
then press ENTER.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN USA
7001 or 7002 Switch System
6430 SourceMeter
Trigger
Link
Trigger
Link Cable
(8501)
Trigger
Link
OUT
IN
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
INPUT/
OUTPUT
42V
PEAK
250V
PEAK
TRIGGER
LINK
4-WIRE
SENSE
HI
LO
LINE RATING
100-240VAC
50, 60, HZ
100VA MAX
RS232
IEEE-488
(ENTER IEEE ADDRESS
WITH FRONT PANEL MENU)
250V
PEAK
5V
PEAK
5V
PEAK
5V
PK
V,
Ω
GUARD
GUARD
SENSE
LINE FUSE
SLOWBLOW
2.5A, 250V
INTERLOCK-
DIGITAL I/O
REMOTE
PreAmp
!
Figure 10-6
Trigger link
connections
Содержание 6430
Страница 26: ......
Страница 32: ......
Страница 78: ...2 14 Connections ...
Страница 98: ...3 20 Basic Source Measure Operation ...
Страница 138: ...5 30 Source Measure Concepts ...
Страница 156: ...6 18 Range Digits Speed and Filters ...
Страница 168: ...7 12 Relative and Math ...
Страница 176: ...8 8 Data Store ...
Страница 202: ...9 26 Sweep Operation ...
Страница 248: ...11 22 Limit Testing ...
Страница 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Страница 418: ...17 108 SCPI Command Reference ...
Страница 450: ...18 32 Performance Verification ...
Страница 477: ...A Specifications ...
Страница 489: ...B StatusandErrorMessages ...
Страница 498: ...B 10 Status and Error Messages ...
Страница 499: ...C DataFlow ...
Страница 503: ...D IEEE 488BusOverview ...
Страница 518: ...D 16 IEEE 488 Bus Overview ...
Страница 519: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 523: ...F MeasurementConsiderations ...
Страница 539: ...G GPIB488 1Protocol ...
Страница 557: ......