17-40
SCPI Command Reference
CALCulate3
Provides statistical data on buffer readings
Select statistic
:FORMat <name>
:CALCulate3:FORMat <name>
Specify CALC3 format
Parameters
<name> =
MEAN
Mean value of readings in buffer
SDEViation
Standard deviation of readings in buffer
MAXimum
Largest reading in buffer
MINimum
Lowest reading in buffer
PKPK
MAXimum - MINimum
Query
:FORMat?
Query programmed math format
Description
This command is used to select the desired statistic on readings stored in the
buffer. See Section 8 for details on these statistics.
Readings stored in the buffer can be
“
raw
”
measured readings, the results of
the CALC1 calculation, or CALC2 readings. The :TRACe:FEED command
in the :TRACe Subsystem is used to select the type of readings to store.
Acquire statistic
:DATA?
:CALCulate3:DATA?
Read CALC3 result
Description
This query command is used to perform the selected statistic operation and
read the result(s). The result(s) is always returned in ASCII format.
If the buffer is con
fi
gured to store
“
raw
”
measured readings (:TRACe:FEED
SENSe1) and multiple functions were measured, the selected statistic opera-
tion will be performed on all the measured readings. For example, if voltage
and current measurements were stored in the buffer, then the selected statis-
tic operation will be performed on both readings. Statistics for multiple
measurement functions are returned in the following order:
voltage statistic, current statistic, resistance statistic
Statistic operations are not performed on TIME and STATus data elements
that are stored in the buffer.
If the buffer is con
fi
gured to store the result of CALC1 or CALC2
(:TRACe:FEED CALC1 or CALC2), only one result will be returned by
this query command.
Notes:
1. If there is no data in the buffer, the NAN (not a number) value +9.91e37
will be returned.
Содержание 6430
Страница 26: ......
Страница 32: ......
Страница 78: ...2 14 Connections ...
Страница 98: ...3 20 Basic Source Measure Operation ...
Страница 138: ...5 30 Source Measure Concepts ...
Страница 156: ...6 18 Range Digits Speed and Filters ...
Страница 168: ...7 12 Relative and Math ...
Страница 176: ...8 8 Data Store ...
Страница 202: ...9 26 Sweep Operation ...
Страница 248: ...11 22 Limit Testing ...
Страница 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Страница 418: ...17 108 SCPI Command Reference ...
Страница 450: ...18 32 Performance Verification ...
Страница 477: ...A Specifications ...
Страница 489: ...B StatusandErrorMessages ...
Страница 498: ...B 10 Status and Error Messages ...
Страница 499: ...C DataFlow ...
Страница 503: ...D IEEE 488BusOverview ...
Страница 518: ...D 16 IEEE 488 Bus Overview ...
Страница 519: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 523: ...F MeasurementConsiderations ...
Страница 539: ...G GPIB488 1Protocol ...
Страница 557: ......