88 Chapter 8: ChartView & ChartView Plus
TempScan / MultiScan User's Manual
Note:
When using a channel as a reference channel, remember that lower numbered channels
are calculated ahead of higher numbered channels, e.g., CH1 is calculated ahead of CH2.
This makes a difference when subtracting a reference channel in regard to the time aspect
of the reference value.
If the channel is configured for Volts,
the units automatically change to V; at this point you have
the option of selecting the millivolt unit (mV). The user can also change each channel to apply a
linear equation (
mX+b
) to the data. When the
mX+b
option is selected the Configure Engineering
Units dialog box is displayed. You are prompted to define “
m
” and “
b
” and the engineering units
label; as well as which standard units the
mX+b
equation should be applied to. The engineering
units chosen will be displayed in the “Units” column, and the “
mX+b
” equation will be applied to
the reading from the device before the reading is displayed or written to disk.
The
mX+b
option is convenient for obtaining a value which has a linear relation to a channel
reading.
X
is the scanned voltage value read back from the acquisition device
m
is [but not limited to] a
proportionality constant
or
gain factor,
but not zero
.
b
is the
offset value
(the value of the calculated reading when the scanned value is “zero”).
Alarm Configuration Columns
1.
Low Limit
Set alarm low limit.
2.
High Limit
Set alarm high limit.
3.
Hysteresis
A hysteresis value can be set for each channel to avoid a “chatter” problem.
4.
Alarm #
You can assign one alarm number (1 to 32) to a channel. Any number of channels
can be assigned to the same alarm number. A given digital output will be set “true” (1) if at least
one of the common channels (assigned to a single alarm) enters an alarm state.
Acquisition Setup Dialog Box
The information entered in the Acquisition Setup dialog box is used by the Arm Acquisition command
to set up the acquisition of data to disk. When the trigger is satisfied, the scans are collected at the
selected scan frequency and stored to disk in the designated file.
The Acquisition Setup dialog box can be accessed in one of three ways: a) clicking on the Channel
Configuration button (item 8 in Main Window figure), b) clicking on the Acquisition Setup tab from
the Channels or Data Destination dialog boxes, c) using the Setup pull-down menu and selecting
Acquisition
.
Содержание MultiScan 1200
Страница 2: ...TempScan MultiScan User s Manual...
Страница 4: ...TempScan MultiScan User s Manual...
Страница 10: ...vi TempScan MultiScan User s Manual Notes...
Страница 20: ...10 Chapter 1 System Overview TempScan MultiScan User s Manual Notes...
Страница 106: ...96 Chapter 8 ChartView ChartView Plus TempScan MultiScan User s Manual Notes...
Страница 130: ...120 Chapter 9 TempView TempScan MultiScan User s Manual...
Страница 150: ...140 Chapter 9 TempView TempScan MultiScan User s Manual Notes...
Страница 166: ...156 Chapter 10 PostView TempScan MultiScan User s Manual Notes...
Страница 234: ...224 Chapter 13 System Operation TempScan MultiScan User s Manual Notes...
Страница 248: ...238 Chapter 14 System Calibration TempScan MultiScan User s Manual Notes...
Страница 284: ...274 Chapter 15 Program Examples TempScan MultiScan User s Manual Notes...
Страница 330: ...320 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Страница 342: ...332 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Страница 358: ...348 Appendix TempScan MultiScan User s Manual Notes...
Страница 364: ...354 Index TempScan MultiScan User s Manual Notes...