![IOtech MultiScan 1200 Скачать руководство пользователя страница 212](http://html1.mh-extra.com/html/iotech/multiscan-1200/multiscan-1200_user-manual_2088024212.webp)
202 Chapter 13: System Operation
TempScan / MultiScan User's Manual
Buffer
Buffer
N/A
Read Pointer >
Scan -100
N/A
Scan -099
N/A
Scan -098
…
Trigger Block
1 Pre-Trigger
Scans
…
Trigger Block
1 Pre-Trigger
Scans
Trigger Scan >
N/A
Trigger Scan >
Scan 000
N/A
Scan 001
…
…
N/A
Write Pointer >
Scan 050
…
…
Stop Scan >
N/A
Trigger Block
1 Post-Trigger
Scans
Stop Scan >
N/A
Trigger Block
1 Post-Trigger
Scans
N/A
N/A
…
…
N/A
N/A
…
…
End Scan >
N/A
Trigger Block
1 Post-Stop
Scans
End Scan >
N/A
Trigger Block
1 Post-Stop
Scans
(1) The Acquisition Buffer is empty.
(2) The Stop event has not yet occurred.
Part 1 – The Acquisition Buffer is empty.
The following describes a Query the Buffer Status String (
U6
) response
when the Acquisition Buffer is empty. The Time/Data stamping is in absolute ASCII format.
0000000,0000000,-0999999,00:00:00.000,00/00/00,-0999999,00:00:00.000,
00/00/00,-0999999,00
Field
Field Name
Buffer Status String Response
Response Description
1
Number of Trigger Blocks Available
0000000
Zero
2
Number of Scans Available
0000000
Zero
3
Current Position of Read Pointer
-0999999
Undefined
4
Time/Date Stamping of Trigger Event
00:00:00.000,00/00/00
Absolute ASCII format
5
Position of Stop Event Pointer
-0999999
Undefined
6
Time/Date Stamping of Stop Event
00:00:00.000,00/00/00
Absolute ASCII format
7
Position of End Scan Pointer
-0999999
Undefined
8
Status of Current Trigger Block
00
Not being acquired
Part 2 – The Stop event has not yet occurred.
The following describes a
U6
response when the
Trigger
event
has occurred but the
Stop
event has not yet occurred. The
Current Position of Read
Pointer
field has the value of
the first available
Pre-Trigger
scan -100, and an additional 50
Post-Trigger
scans are available. The Time/Data
stamping is in absolute ASCII format.
0000001,0000151,-0000100,12:01:43.100,08/29/96,-0999999,00:00:00.000,
00/00/00,-0999999,00
Field
Field Name
Buffer Status String Response
Response Description
1
Number of Trigger Blocks Available
0000001
One
2
Number of Scans Available
0000151
100 + 1 + 50
3
Current Position of Read Pointer
-0000100
Pre-Trigger Scan -100
4
Time/Date Stamping of Trigger Event
12:01:43.100,08/29/96
Absolute ASCII format
5
Position of Stop Event Pointer
-0999999
Undefined
6
Time/Date Stamping of Stop Event
00:00:00.000,00/00/00
Absolute ASCII format
7
Position of End Scan Pointer
-0999999
Undefined
8
Status of Current Trigger Block
00
Still being acquired
Query (U6) Buffer Operation – Single Trigger Block, Parts 1 & 2 (of 4)
Содержание MultiScan 1200
Страница 2: ...TempScan MultiScan User s Manual...
Страница 4: ...TempScan MultiScan User s Manual...
Страница 10: ...vi TempScan MultiScan User s Manual Notes...
Страница 20: ...10 Chapter 1 System Overview TempScan MultiScan User s Manual Notes...
Страница 106: ...96 Chapter 8 ChartView ChartView Plus TempScan MultiScan User s Manual Notes...
Страница 130: ...120 Chapter 9 TempView TempScan MultiScan User s Manual...
Страница 150: ...140 Chapter 9 TempView TempScan MultiScan User s Manual Notes...
Страница 166: ...156 Chapter 10 PostView TempScan MultiScan User s Manual Notes...
Страница 234: ...224 Chapter 13 System Operation TempScan MultiScan User s Manual Notes...
Страница 248: ...238 Chapter 14 System Calibration TempScan MultiScan User s Manual Notes...
Страница 284: ...274 Chapter 15 Program Examples TempScan MultiScan User s Manual Notes...
Страница 330: ...320 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Страница 342: ...332 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Страница 358: ...348 Appendix TempScan MultiScan User s Manual Notes...
Страница 364: ...354 Index TempScan MultiScan User s Manual Notes...