TempScan / MultiScan User's Manual
Chapter 2: TempScan/1100 & MultiScan/1200 15
Data Storage & Format
Storage:
128 K reading (256 Kbyte) standard; optional 500 K reading (1 Mbyte), 2 M reading (4
Mbyte), 4 M reading (8 Mbyte).
Data Formats:
ASCII and binary; binary format returns a 16-bit compensated and linearized
temperature value (0.1°C/bit); user-programmable for hi/low byte or low/hi byte. Note that high
speed DMA transfers are binary format only.
Statistical Parameters:
High, Low, and Last available per channel.
For MultiScan/1200:
Not
available when in single-channel burst mode.
Time Stamping:
Available for each scan group and for each channel’s high, low, and last parameters.
For MultiScan/1200:
Not available when in single-channel burst mode.
Time Format:
Absolute Time/Date stamping (
hh:mm:ss.mil,MM/DD/YY
), relative Time/Date
stamping (
+hh:mm:ss.mil,DDDDDDD
) and scan interval timebase (
hh:mm:ss.t
).
For
MultiScan/1200:
Not available when in single-channel burst mode.
Alarm Stamp:
Available for each scan group.
For MultiScan/1200:
Not available when in single-
channel burst mode.
Digital Input Stamp:
Available for each scan group.
For MultiScan/1200:
Not available when in
single-channel burst mode.
IEEE 488 Interface
CAUTION
CAUTION
CAUTION
CAUTION
The IEEE 488 terminal must only be used to control a non-isolated IEEE 488
system. The common mode voltage (cable shell to earth) must be zero.
Interface Use:
Digital communication (as opposed to analog) for IEEE 488 compliant computer
platforms, as well as IEEE 488 compliant
platform-independent
configurations. Messages sent 1
byte (8 bits) at a time. Supports data rates up to 1 Mbyte/sec. Up to 15 devices can be connected
to one bus. Total bus length up to 20 meters. Allowable cable distance between devices is up to 2
meters. Message transactions are hardware handshaked.
Installation Category:
For CE:
Category 1.
Implementation:
SH1, AH1, T6, TE4, L4, LE4, SR1, PP0, RL0, DC1, DT1, C0, E1.
Programmable Parameters:
Alarm set points, thermocouple type, temperature units, Trigger level,
Pre-Trigger and Post-Trigger scan interval, Trigger mode, SRQ mask, scan count, Pre-Trigger
count, digital input, digital output, real time settings, data output format, and terminators.
Data Transfer Speed:
Up to >300 Kbytes/s.
Connector:
Standard IEEE 488 connector with metric studs.
RS-232/RS-422 Serial Interface
CAUTION
CAUTION
CAUTION
CAUTION
The RS-232/RS-422 terminal is only for connecting devices having signals at serial
communications levels.
Installation Category:
For CE:
Category 1.
Baud Rates:
300, 600, 1200, 2400, 4800 and 9600.
Data Bits:
8.
Stop Bits:
1.
Parity:
Even, Odd, None.
Handshaking:
RTS/CTS, XON/XOFF (for ASCII transmissions only).
Connector:
Male DB-9.
Содержание MultiScan 1200
Страница 2: ...TempScan MultiScan User s Manual...
Страница 4: ...TempScan MultiScan User s Manual...
Страница 10: ...vi TempScan MultiScan User s Manual Notes...
Страница 20: ...10 Chapter 1 System Overview TempScan MultiScan User s Manual Notes...
Страница 106: ...96 Chapter 8 ChartView ChartView Plus TempScan MultiScan User s Manual Notes...
Страница 130: ...120 Chapter 9 TempView TempScan MultiScan User s Manual...
Страница 150: ...140 Chapter 9 TempView TempScan MultiScan User s Manual Notes...
Страница 166: ...156 Chapter 10 PostView TempScan MultiScan User s Manual Notes...
Страница 234: ...224 Chapter 13 System Operation TempScan MultiScan User s Manual Notes...
Страница 248: ...238 Chapter 14 System Calibration TempScan MultiScan User s Manual Notes...
Страница 284: ...274 Chapter 15 Program Examples TempScan MultiScan User s Manual Notes...
Страница 330: ...320 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Страница 342: ...332 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Страница 358: ...348 Appendix TempScan MultiScan User s Manual Notes...
Страница 364: ...354 Index TempScan MultiScan User s Manual Notes...