180 Chapter 12: System Configuration
TempScan / MultiScan User's Manual
To enable the gap-free acquisition with two timebases, you must perform the following:
•
Specify the two trigger events – the
Trigger
(trigger start event) and the
Stop
(trigger stop event) –
via the
Set Trigger Configuration
(
T
) command.
•
Set the two timebase intervals – the
normal
and the
acquisition
scan intervals – via the
Set Scan
Interval
(
I
) command.
•
Set the
Pre-Trigger
count to
-1
, via the
pre
parameter of the
Set Counts
(
Y
) command. If more
than one transition cycle is desired, enable the
Auto Re-arm
flag of the
Set Trigger Configuration
(
T
) command to allow each transition cycle to be stored in its own
Trigger Block
within the buffer.
Using this method, continuous data may be collected as long as the application program can read data
out of the Acquisition Buffer before the buffer overruns. If the buffer overruns, the data read may not
be continuous.
Trigger Configuration
Trigger Start & Stop Events
The programmed system events are the driving force behind any acquisition. For an acquisition to take
place, the
Set Trigger Configuration
(
T
) command is used as the central command to assign the
following events to the TempScan/1100 or MultiScan/1200 unit: The
Trigger
(trigger start event) and
the
Stop
(trigger stop event). The
Trigger
and
Stop
may also be assigned different trigger sources. In
addition, the
Set Trigger Configuration
(
T
) command also determines whether or not, after the initial
acquisition, the acquisition will be re-enabled automatically, and whether or not the
Trigger
(trigger
start event) should be synchronized with the
Pre-Trigger
scan interval.
Upon the execution of this command, the unit will enter acquisition mode, and the
TRIGGER
LED
indicator on the front panel will flash. If
Pre-Trigger
scans have been configured via the
Set Counts
(
Y
) command, then the sampling for
Pre-Trigger
data will begin at this time. Specifically, the
Set
Trigger Configuration
(
T
) command is used to set the following parameters:
•
Trigger (trigger start event):
This parameter is the event that is to take place in order for the unit
to begin acquiring scans. When the unit detects a
Trigger
, it will stop acquiring
Pre-Trigger
scans
(if configured), and begin acquiring
Post-Trigger
scans at that point. The first of these
Post-
Trigger
scans, the
Trigger
scan, will be Time/Date stamped for later reference, such as in a
response to the following
User Status
(
U
) command query:
U6
– Query the Buffer Status String.
Also, when the
Trigger
is detected, the Triggered Bit (Bit 1) of the Status Byte Register (
STB
) will
be set and the
TRIGGER
LED indicator on the front panel will turn on.
•
Stop (trigger stop event):
This parameter is the event that is to take place in order for the unit to
stop acquiring scans. When the unit detects a
Stop
, it will stop acquiring
Post-Trigger
scans at that
point, unless
Post-Stop
scans are defined via the
Set Counts
(
Y
) command. The last of these
Post-
Trigger
scans, the
Stop
scan, will be Time/Date stamped for later reference, such as in a response
to the following
User Status
(
U
) command query:
U6
– Query the Buffer Status String. Also, when
the
Stop
is recognized, the Triggered Bit (Bit 1) of the Status Byte Register (
STB
) will be cleared
and the
TRIGGER
LED indicator on the front panel will turn off.
Содержание MultiScan 1200
Страница 2: ...TempScan MultiScan User s Manual...
Страница 4: ...TempScan MultiScan User s Manual...
Страница 10: ...vi TempScan MultiScan User s Manual Notes...
Страница 20: ...10 Chapter 1 System Overview TempScan MultiScan User s Manual Notes...
Страница 106: ...96 Chapter 8 ChartView ChartView Plus TempScan MultiScan User s Manual Notes...
Страница 130: ...120 Chapter 9 TempView TempScan MultiScan User s Manual...
Страница 150: ...140 Chapter 9 TempView TempScan MultiScan User s Manual Notes...
Страница 166: ...156 Chapter 10 PostView TempScan MultiScan User s Manual Notes...
Страница 234: ...224 Chapter 13 System Operation TempScan MultiScan User s Manual Notes...
Страница 248: ...238 Chapter 14 System Calibration TempScan MultiScan User s Manual Notes...
Страница 284: ...274 Chapter 15 Program Examples TempScan MultiScan User s Manual Notes...
Страница 330: ...320 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Страница 342: ...332 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Страница 358: ...348 Appendix TempScan MultiScan User s Manual Notes...
Страница 364: ...354 Index TempScan MultiScan User s Manual Notes...