TempScan / MultiScan User's Manual
Chapter 16: API Command Reference 333
*
U15
: Query the product information. This request responds with the current
TempScan/1100 or MultiScan/1200 product information including version levels.
*
U16
: Query the measuring mode parameters (MultiScan/1200 only). For the
MultiScan/1200 only, this request is equivalent to the following string of measuring mode
queries:
M#?F#?W#?
Consequently, this request responds with the following measuring
mode data in their respective forms:
M#mode F#freq W#wt
*
U17
: Query the Root Mean Square (RMS) value (MultiScan/1200 only). For
the MultiScan/1200 only, this request responds with the Root Mean Square (RMS) value
computed from the last-completed acquisition in single-channel high-speed burst mode. For
more information, see section
Measuring Modes (MultiScan/1200 Only)
in the chapter
System Configuration
.
*
U18
: Query the acquisition states and various system flags. This request
responds with an integer
nnn
of the form
000
≤≤≤≤
nnn
≤≤≤≤
255
, where the separate values are as follows,
and multiple values are summed. The
U18
Bits 7 through 4 respond with the following individual values:
128
- Acquisition Block Available (Bit 7)
032
- Not used.
064
- Power up under default configuration
(Bit 6). Otherwise, power up under the
last configuration.
016
- Scan Alarm Stamping Enabled (Bit 4)
Meanwhile, the
U18
Bits 3 and 2 respond with the following combined values:
012
- Not used.
004
- Absolute Scan Time Stamping Enabled
(Bit 2 only)
008
- Relative Scan Time Stamping
Enabled (Bit 3 only)
000
- Scan Time Stamping Not Enabled
(Neither Bits 3 nor 2)
Lastly, the
U18
Bits 1 and 0 respond with the following combined values:
003
- Acquiring Post-Stop Scan Data
(Both Bits 1 and 0)
001
- Armed – Acquisition is armed and may be
acquiring Pre-Trigger Scan Data (Bit 0 only)
002
- Acquiring Post-Trigger Scan Data
(Bit 1 only)
000
- Idle – No acquisition is armed or acquiring any
scan data (Neither Bit 1 nor Bit 0)
Note:
Except for calibration errors, any other error conditions are cleared after the status string is read by the
controller. Meanwhile, calibration errors are cleared when the unit is calibrated.
Note:
Status strings are returned when the unit is next addressed to TALK. If the unit is configured to Trigger on
TALK, requesting a status report will not cause the unit to trigger.
EXAMPLES
Refer to the EXAMPLES section for each of the following commands:
*K -
Change Calibration Keyword
C -
Configure Channels
C# -
Select Card
F -
Set Data Format
K -
Enter Calibration Mode
Q -
Set Query Terminator
X -
Execute
Y -
Set Counts
Содержание MultiScan 1200
Страница 2: ...TempScan MultiScan User s Manual...
Страница 4: ...TempScan MultiScan User s Manual...
Страница 10: ...vi TempScan MultiScan User s Manual Notes...
Страница 20: ...10 Chapter 1 System Overview TempScan MultiScan User s Manual Notes...
Страница 106: ...96 Chapter 8 ChartView ChartView Plus TempScan MultiScan User s Manual Notes...
Страница 130: ...120 Chapter 9 TempView TempScan MultiScan User s Manual...
Страница 150: ...140 Chapter 9 TempView TempScan MultiScan User s Manual Notes...
Страница 166: ...156 Chapter 10 PostView TempScan MultiScan User s Manual Notes...
Страница 234: ...224 Chapter 13 System Operation TempScan MultiScan User s Manual Notes...
Страница 248: ...238 Chapter 14 System Calibration TempScan MultiScan User s Manual Notes...
Страница 284: ...274 Chapter 15 Program Examples TempScan MultiScan User s Manual Notes...
Страница 330: ...320 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Страница 342: ...332 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Страница 358: ...348 Appendix TempScan MultiScan User s Manual Notes...
Страница 364: ...354 Index TempScan MultiScan User s Manual Notes...