TempScan / MultiScan User's Manual
Chapter 15: Program Examples 263
The following program is
BIN_ACQ.BAS
in the
\EXAMPLES\
subdirectory. It will configure 12
thermocouple and 12 volts channels, set up the MultiScan/1200 for acquisition, and then after the
acquisition is complete, collect the buffered channel readings in a Binary format rather than the default
ASCII format. The Binary format is inherently faster than ASCII format, but the data must be
deciphered after it is collected in order to yield the temperature values.
•
Lines 1-3:
Although not necessary, it is good practice to reset the MultiScan/1200 at the start of
your application by sending the
Reset
Power-On
(
*R
) command. Wait a few seconds.
•
Lines 4-8:
To be certain that the reset was successful and the device is ready, Serial Poll the device
until the proper status is returned.
•
Lines 9-10:
The first thing necessary to start scanning is to setup the measuring mode. In this case,
the MultiScan/1200 will be setup for line-cycle integration / high-speed multi-channel mode.
•
Lines 11-12:
The
Configure Channels
(
C
) command will be used once to configure a range of
channels from 1 to 12 as Type T thermocouples, and used again to configure a range of channels
from 25 to 36 as 10-Volt DC inputs.
•
Line 13:
Using the
Set Counts
(
Y
) command, 400 scans will be taken while in the
Pre-Trigger
state, 100 while in the
Post-Trigger
state, and 200 while in the
Post-Stop
state. The
Post-Trigger
count is only valid when the
Stop
(trigger stop event) is set to Count via the
Set Trigger
Configuration
(
T
) command. This example will in fact use Count, so this parameter is valid.
•
Line 14:
Using the
Set Scan Interval
(
I
) command, the
Post-Trigger
,
Pre-Trigger
and
Post-Stop
scan intervals are set to the same value of 0.0 second, which defaults the MultiScan/1200
acquisition to fast mode.
•
Line 15:
Using the
Set Trigger Configuration
(
T
) command, the
Trigger
(trigger start event) is
assigned the
Trigger On
(
@
) command character as sent by the controller, while the
Stop
(trigger
stop event)
is assigned to Count. Both the
Auto Re-arm
flag and the synchronization flag are
disabled, as described in previous examples. When the MultiScan/1200 encounters the
Execute
(
X
) command, it will be armed and will start the collection of the
Pre-Trigger
data.
•
Lines 16-25:
After the last
Execute
(
X
) command has been received, the MultiScan/1200 arms
itself and then waits for the
Trigger
(trigger start event). The IEEE 488 Serial Poll (
SPOLL
)
command is used to query the acquisition status of the MultiScan/1200. A
SPOLL
value of 2
signifies that the MultiScan/1200 has detected the
Trigger
.
•
Lines 26-31:
Using the
User Status
(
U
) command
U0
to query the Event Status Register (
ESR
), we
now wait until the Acquisition Complete event is reached.
•
Line 32:
Before reading the data, the
Set Data Format
(
F
) command is sent, instructing the
MultiScan/1200 to return the data in Binary format where every value will be 2 bytes in length.
This command can be issued anytime after the initial reset, and stays in effect until the unit is reset
again or another
F
command is issued.
•
Lines 33-40:
During the acquisition or after the acquisition has been completed, the Acquisition
Buffer can be queried for the data that is available for transfer. The
User Status
(
U
) command
U6
can be used to query the Buffer Status String, as described in previous examples. This example
uses the last parameter of the return string as an indicator of how much data is available to transfer
to the IEEE 488 controller. The function
MID$
extracts 7 characters from the string
U$
starting at
character 9.
•
Lines 41-50:
After calculating the number of scans, an integer-type buffer is then prepared for the
data. The offset and segment pointers are supplied to the IEEE 488 driver so that the data can be
transferred using DMA (Direct Memory Access). The
Read Buffered Data
(
R
) command
R2
instructs the MultiScan/1200 to supply the oldest
Trigger Block
of data that is currently available.
•
Lines 51-52:
At the completion of the DMA transfer, the integer data is now in the buffer
BUFFER%
. To convert the integer data into temperatures and into voltages, simple calculations are
performed.
Содержание MultiScan 1200
Страница 2: ...TempScan MultiScan User s Manual...
Страница 4: ...TempScan MultiScan User s Manual...
Страница 10: ...vi TempScan MultiScan User s Manual Notes...
Страница 20: ...10 Chapter 1 System Overview TempScan MultiScan User s Manual Notes...
Страница 106: ...96 Chapter 8 ChartView ChartView Plus TempScan MultiScan User s Manual Notes...
Страница 130: ...120 Chapter 9 TempView TempScan MultiScan User s Manual...
Страница 150: ...140 Chapter 9 TempView TempScan MultiScan User s Manual Notes...
Страница 166: ...156 Chapter 10 PostView TempScan MultiScan User s Manual Notes...
Страница 234: ...224 Chapter 13 System Operation TempScan MultiScan User s Manual Notes...
Страница 248: ...238 Chapter 14 System Calibration TempScan MultiScan User s Manual Notes...
Страница 284: ...274 Chapter 15 Program Examples TempScan MultiScan User s Manual Notes...
Страница 330: ...320 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Страница 342: ...332 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Страница 358: ...348 Appendix TempScan MultiScan User s Manual Notes...
Страница 364: ...354 Index TempScan MultiScan User s Manual Notes...