26 Chapter 2: TempScan/1100 & MultiScan/1200
TempScan / MultiScan User's Manual
Scanning Card & Channel Expansion
Scanning Card Expansion
Each TempScan/1100 or MultiScan/1200 unit can accept one scanning card to provide signal
conditioning. However, if your application demands more channels then you can expand the master
unit’s capabilities with either or both types of expansion units: Exp/10A and/or Exp/11A. Each
Exp/10A expansion unit allows for the addition of two scanning cards, while each Exp/11A expansion
unit allows for the addition of ten scanning cards. Up to fifteen Exp/10A units can be attached to the
master unit, giving a maximum of 30 additional scanning-card slots. Similarly, up to three Exp/11A
units can be linked to the master unit, also giving a maximum of 30 additional scanning-card slots, as
shown in the following table.
Expansion Capabilities
Exp/10A
Exp/11A
Number of expansion scanning-card slots per expansion unit
2
10
Number of identical expansion units that can be linked to the master unit
15
3
Maximum number of expansion scanning-card slots
30
30
Note:
If a combination of Exp/10A and Exp/11A expansion units are linked together, then the
maximum number of expansion scanning-card slots is still 30.
Channel Expansion
The TempScan/1100 system can be expanded up to 992 channels, while the MultiScan/1200 system
can be expanded up to 744 channels. This is easily accomplished via a
master/slave architecture
wherein a main or
master
unit can be connected to as many expansion
slave
units as allowable within
the maximum of 30 scanning cards.
Each Exp/10A or Exp/11A expansion unit has a form factor identical to that of either the
TempScan/1100 or MultiScan/1200 master unit, allowing the expansion unit to accept the same
scanning cards as its master unit. When connected to the TempScan/1100, the Exp/10A or Exp/11A is
configurable for 32 or 64 input channels, providing a total expansion capacity of up to 992 channels.
When connected to the MultiScan/1200, the Exp/10A or Exp/11A is configurable for 24 or 48 input
channels, providing a total expansion capacity of up to 744 channels.
An expansion unit can only be controlled by the TempScan/1100 or MultiScan/1200 master unit, and
no digital I/O or alarms are included. However, expansion channels can be programmed to stimulate
the alarms located on the master unit. From the programmer’s perspective, channels are accessed in the
same way as channels in the master unit. When the master unit detects the presence of expansion units
during its power-on sequence, it makes the additional channels available to the programmer. The
U8
command is available to query the master unit for the total number of channels in the system.
A general description of both types of expansion units, their connections and their configurations, is
discussed in the following chapters.
Содержание MultiScan 1200
Страница 2: ...TempScan MultiScan User s Manual...
Страница 4: ...TempScan MultiScan User s Manual...
Страница 10: ...vi TempScan MultiScan User s Manual Notes...
Страница 20: ...10 Chapter 1 System Overview TempScan MultiScan User s Manual Notes...
Страница 106: ...96 Chapter 8 ChartView ChartView Plus TempScan MultiScan User s Manual Notes...
Страница 130: ...120 Chapter 9 TempView TempScan MultiScan User s Manual...
Страница 150: ...140 Chapter 9 TempView TempScan MultiScan User s Manual Notes...
Страница 166: ...156 Chapter 10 PostView TempScan MultiScan User s Manual Notes...
Страница 234: ...224 Chapter 13 System Operation TempScan MultiScan User s Manual Notes...
Страница 248: ...238 Chapter 14 System Calibration TempScan MultiScan User s Manual Notes...
Страница 284: ...274 Chapter 15 Program Examples TempScan MultiScan User s Manual Notes...
Страница 330: ...320 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Страница 342: ...332 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Страница 358: ...348 Appendix TempScan MultiScan User s Manual Notes...
Страница 364: ...354 Index TempScan MultiScan User s Manual Notes...