TempScan / MultiScan User's Manual
Chapter 14: System Calibration 237
Calibration of RTD Scanning Cards
Note:
Prior to each calibration session, disable calibration protection by setting the rear panel DIP
microswitch 9 to the up (
1
) position. Then enable
Calibration Mode
via the
Enter Calibration
Mode
(
K
) command with a valid password. If the default calibration password has not been
changed, enter
K12345
.
Note:
At the end of each calibration session, disable
Calibration Mode
with the
End Calibration
Mode
(
E
) command. Then re-enable calibration protection by resetting DIP microswitch 9 to
the down (
0
) position.
To Calibrate an RTD Card for Offset
1.
Connect a Copper Short to Channel 1 (across R1-, S1-, S1+, R1+) on the scanning card being
calibrated. Use standard copper hookup wire.
Do not use thermocouple wire.
Slide the scanning
card into the previously-calibrated master chassis.
2.
Power up the unit and wait at least one hour before continuing.
3.
Send the
Calibrate Channel Offset
(
H
) command
Hchan
– where
chan
refers to the first channel
number on the scanning card being calibrated (refer to the previous table for valid offset channel
numbers) – and wait for the command to complete. The
TRIGGER
LED indicator will flash for a
few seconds.
4.
Monitor the Calibration Status Register (
CSR
) to ensure that the
Calibrate Channel Offset
(
H
)
command completed without error. Refer to the
User Status
(
U
) command
U2
to monitor and
decode the Calibration Status Register (
CSR
).
To Calibrate an RTD Card for Gain
CAUTION
CAUTION
CAUTION
CAUTION
Ensure that the RTD scanning card has been calibrated for offset. If not, follow
the RTD card offset calibration procedure as described in the previous text.
1.
Connect a precision 3-wire or 4-wire resistance source of
xxx.xx
ohms to Channel 1 on the
scanning card being calibrated. Slide the scanning card into the previously-calibrated master
chassis.
2.
Power up the unit and wait at least one hour before continuing.
3.
Enter the
Set RTD Gain Calibration Reference
(
*G
) command
*Ggain
where
gain
is the
appropriate resistance value
nnn.nn
in ohms.
4.
Send the following
Calibrate Channel Gain
(
G
) command
Gchan,type
– where
chan
refers to
the first channel number on the scanning card being calibrated (refer to the previous table for valid
gain channel numbers):
•
3-Wire RTD:
Gchan,16
•
4-Wire RTD:
Gchan,17
Wait for the command to complete. The
TRIGGER
LED indicator will flash for a few seconds.
5.
Monitor the Calibration Status Register (
CSR
) to ensure that the
Calibrate Channel Gain
(
G
)
command completed without error. Refer to the
User Status
(
U
) command
U2
to monitor and
decode the Calibration Status Register (
CSR
).
Содержание MultiScan 1200
Страница 2: ...TempScan MultiScan User s Manual...
Страница 4: ...TempScan MultiScan User s Manual...
Страница 10: ...vi TempScan MultiScan User s Manual Notes...
Страница 20: ...10 Chapter 1 System Overview TempScan MultiScan User s Manual Notes...
Страница 106: ...96 Chapter 8 ChartView ChartView Plus TempScan MultiScan User s Manual Notes...
Страница 130: ...120 Chapter 9 TempView TempScan MultiScan User s Manual...
Страница 150: ...140 Chapter 9 TempView TempScan MultiScan User s Manual Notes...
Страница 166: ...156 Chapter 10 PostView TempScan MultiScan User s Manual Notes...
Страница 234: ...224 Chapter 13 System Operation TempScan MultiScan User s Manual Notes...
Страница 248: ...238 Chapter 14 System Calibration TempScan MultiScan User s Manual Notes...
Страница 284: ...274 Chapter 15 Program Examples TempScan MultiScan User s Manual Notes...
Страница 330: ...320 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Страница 342: ...332 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Страница 358: ...348 Appendix TempScan MultiScan User s Manual Notes...
Страница 364: ...354 Index TempScan MultiScan User s Manual Notes...