196 Chapter 12: System Configuration
TempScan / MultiScan User's Manual
Power-Up Configuration
The TempScan/1100 or MultiScan/1200 unit has an internally-stored factory-default configuration
which can be recalled at any time. You can program the unit to power-up in a user-defined state, or if
preferred, you can configure the unit to power-up with the default configuration.
During normal operations, the factory defaults may be recalled by issuing the
Restore Factory Defaults
(
*F
) command. When this command is processed, the TempScan/1100 or MultiScan/1200 will load
the factory default configuration and will operate accordingly. However, this command may not be
issued while an acquisition is currently configured.
The TempScan/1100 or MultiScan/1200 unit is normally configured to power up with the last known
configuration at the time that it was powered down, or at the time that the
Reset Power-On
(
*R
)
command is issued. However, the unit may be configured to always power-up either under the factory
default configuration or under the last-known configuration, via the
Set Power-Up Configuration
(
*S
)
command.
Note:
Configuring the TempScan/1100 or MultiScan/1200 master unit to power up under the factory
defaults or recalling the factory default settings will not affect pre-existing calibration factors
of the master unit or its slave units.
The following list describes the execution steps and effects when the TempScan/1100 or
MultiScan/1200 is configured to power up under the factory default configuration:
•
Step 1:
Q7,0,0,0,0
. Using the
Set Query Terminator
(
Q
) command, set the general response
terminator to
CR-LF/EOI
. Set all other terminators to
None
.
•
Step 2:
V0
. Using the
Set User Terminator
(
V
) command, set the user terminator to ASCII
hexadecimal value
00
, or
NUL
.
•
Step 3:
F0,0
. Using the
Set Data Format
(
F
) command, set the data format to Engineering Units,
Degrees Celsius.
•
Step 4:
M0
. Using the
Set SRQ Mask
(
M
) command, set the SRQ Mask to
000
.
•
Step 5:
N0
. Using the
Set Event Mask
(
N
) command, set the Event Status Enable Register (
ESE
) to
000
.
•
Step 6:
*C
. Using the
Clear Channel Configuration
(
*C
) command, clear all channel
configurations.
•
Step 7:
Y0,0,0
. Using the
Set Counts
(
Y
) command, set all acquisition counts to 0.
•
Step 8:
I00:00:00.0,00:00:00.0
. Using the
Set Scan Interval
(
I
) command, set both
timebase intervals to 0 seconds, or
fast mode
.
•
Step 9:
L1,0,0
. Using the
Set Trigger Level
(
L
) command, clear the trigger level configurations.
•
Step 10:
T0,0,0,0
. Using the
Set Trigger Configuration
(
T
) command, clear the acquisition
trigger configurations.
•
Step 11:
M#0
. MultiScan/1200 only. Using the
Set Measuring Mode
(
M#
) command, set the
measuring mode to line-cycle integration/high-speed multi-channel mode.
•
Step 12:
F#20000.0
. MultiScan/1200 only. Using the
Set Burst Mode Frequency
(
F#
)
command, set the burst mode frequency to 20 kHz.
•
Step 13:
W#32
. MultiScan/1200 only. Using the
Set Averaging Weight
(
W#
) command, set the
sampling weight to 32.
For more information, see chapter
Power & Assembly
.
Содержание MultiScan 1200
Страница 2: ...TempScan MultiScan User s Manual...
Страница 4: ...TempScan MultiScan User s Manual...
Страница 10: ...vi TempScan MultiScan User s Manual Notes...
Страница 20: ...10 Chapter 1 System Overview TempScan MultiScan User s Manual Notes...
Страница 106: ...96 Chapter 8 ChartView ChartView Plus TempScan MultiScan User s Manual Notes...
Страница 130: ...120 Chapter 9 TempView TempScan MultiScan User s Manual...
Страница 150: ...140 Chapter 9 TempView TempScan MultiScan User s Manual Notes...
Страница 166: ...156 Chapter 10 PostView TempScan MultiScan User s Manual Notes...
Страница 234: ...224 Chapter 13 System Operation TempScan MultiScan User s Manual Notes...
Страница 248: ...238 Chapter 14 System Calibration TempScan MultiScan User s Manual Notes...
Страница 284: ...274 Chapter 15 Program Examples TempScan MultiScan User s Manual Notes...
Страница 330: ...320 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Страница 342: ...332 Chapter 16 API Command Reference TempScan MultiScan User s Manual...
Страница 358: ...348 Appendix TempScan MultiScan User s Manual Notes...
Страница 364: ...354 Index TempScan MultiScan User s Manual Notes...